FISCHERSCOPE
®
X-RAY
Def.MA
Creating a new Def.MA
15.2 Creating a new Def.MA
The creation of a new Def.MA in the thickness mode (see Table 15-1 on Page 113) is
described here. If you create a new Def.MA in another mode, the steps may differ.
Before you start
The WinFTM extension SUPER is activated.
Procedure
1. In the WinFTM main menu select
Calibrate > Def.MA new... >
<Mode, see
Table 15-1 on Page 113>.
The window
Definition of measuring conditions...
appears.
2. In the field
Label
enter an explicit name for the new Def.MA file.
3. In the field located right above select the high voltage and the primary filter (if present).
The high voltage and the primary filter has to fit your measurement task.
Enter the layer composition of your specimen as follows:
Start with the top layer, continue with all other layers and end with the base material
(substrate material) as the last layer.
4. Click
New Layer
.
The window
Definition of Elements New Layer
appears.
5. In the area
Thickness
select an option.
6. In the area
Composition
select an option.
7. Click
Periodic system
.
Component
mode
For the analysis of specimens, which are not only composed of pure
elements, but also contain chemical compounds (e. g. oxides).
WinFTM offers a list of components, which can be used like ele-
ments.
The result display shows the measurement readings of the complete
components, not only the elements the components are composed
of.
Component
mode with bal-
ance
For the analysis of bulk material (massive specimens), if the speci-
men cannot be defined completely by 100 percent and the specimen
is not only composed of pure elements, but also contain chemical
compounds (e. g. oxides).
The result display shows the measurement readings of the complete
components, not only the elements the components are composed
of.
Typical applications: recycling, RoHS, soil specimens, biological
specimens
Tab. 15-1: Def.MA Modes
Mode
Application
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...