background image

 

Page 5 

 

System 

Software-Info 
Config.-Info 
Statistics 
Password 
Config. load 
Config. save 

5.5 

Menu level Switch points 

The  two  switch  outputs  are  configured  by  four  pa-
rameters respectively. 

For switch point 1 these are: 

SP1 On

SP1 Off,SP1 

Delay

SP1 Function

For  switch  point  2  accordingly: 

SP2  On

SP2 

Off,SP2 Delay

SP2 Function

SP1  On

  specifies  the  activation  point, 

SP1  Off

  the 

deactivation point of switch output 1. The values are 
displayed and set in the valid unit. 

Together, the two parameters determine the switch 
function of switch output 1: 

If 

SP1  Off

<

SP1  On

,  the  output  switches  on,  if  the 

measured  value  exceeds 

SP1On

.  It  is  only 

switched off again if the measured value 

SP1 Off

 is 

undercut (hysteresis function). 

If 

SP1 On

  = 

SP1 Off

, the  output switches on if the 

measured  value  exceeds 

SP1  On 

and  off  if  the 

measured value undercuts 

SP1Off

If 

SP1 Off

>

SP1 ON

, the output switches on, if 

SP1 

On

<  Measured  value  <

SP1  Off

  applies  (window 

function). 

Both parameters can be set independently over the 
entire range. 

SP1  Delay

  allows  the  reaction  of  the  switch  output 

1 to be delayed by between 0 and 100 s. This value 
applies equally for switching on and off. 

 

The  following  applies  for  instruments 

with firmware 2.16 and later: 

The  reaction  of  the  switch  output  is  delayed  by 
0…1800 s. 

SP1  Function

  changes  the  function  of  the  switch 

output. It is possible here to define whether the con-
tact  should  work  as  an  open  contact  (NO)  or  a 
break contact (NC). 

5.6 

Menu level Input 

If  there  are  unsteady  pressure  readings  during  op-
eration, you can use the parameters 

Damping

 and 

Zero-point  window

  to  stabilise  the  reading  (and 

the output signal). 

The  effect  of  parameter 

Damping

  (on  the  reading, 

output  signal  and  switching  points,  if  available,  but 
not on the measuring cell!) corresponds to that of a 
capillary  throttle.  You  can  set  the  response  time  to 
pressure  jumps  in  the  range  0.0  to  100  s.  But  with 
maximum damping, it will take more than 2 minutes 
for  the  reading  to  also  reach  zero  after  a  pressure 
jump from nominal pressure (100 %) to zero! 

In  many  cases,  unsteady  readings  are  not  a  prob-
lem  during  normal  operating  mode,  but  this  is  not 
true  for  the  idle  state,  i.e.  if  zero  (differential)  pres-
sure  is  expected.  In such  situations,  the  parameter 

Zero-point  window 

can  be  applied.  Its  value  de-

fines  a  range  of measurement  values  around  zero, 
for  which  the  measuring  value  is  set  to  zero.  The 
reading  will  not  indicate  zero  any  more,  only  if  the 
pressure exceeds these limits. When reaching dou-
ble  the  value,  the  measuring  pressure  and  reading 
match again. This avoids jumps in the reading. 

It  makes  sense  to  set  the 

Offset

  (zero-point  dis-

placement)  if,  without  differential  pressure  (remove 
measuring  line),  the  display  shows  a  value  that  is 
not zero. Before the offset correction, the zero-point 
window should be set to zero. 

Select the 

Offset correction

 parameter and correct 

the reading using the buttons 

þ

 and 

ÿ

 until  zero  is 

shown.  

 

The  following  applies  for  instruments 

with firmware V2.15 and later: 

The  currently  measured  value  is  shown  when  set-
ting  the  offset.  The  zero  point  window  is  not  active 
when the offset is being set. 

5.7 

Measurement 

The  transmitter  output  signal  initially  depends  on 
the sensed pressure. However, you have the option 
of  adjusting  the  output  signal  to  a  large  extent  to 
suit your requirements.  

The  basic  measuring  range  (indicated  on  the  type 
label)  and  the  type  of  output  signal  (voltage  /  cur-
rent see type label) are not variable. 

The  parameters 

Start  of  measuring  range 

and 

End  of  measuring  range

  initially  define  the  two 

pressures  between  which  the  output  signal  will 
change at all. Both values are adjustable across the 
entire  basic  measuring  range.  The  set  values  also 

Summary of Contents for DE46

Page 1: ...fety instructions the in tended use of the device or the limit values given in the technical specifications can be hazardous or cause harm to persons the environment or the plant itself Fischer Mess u...

Page 2: ...situations but which could be useful 2 Application purpose Display and switching device for differential pres sure of gaseous media The instrument is to be ex clusively used for the applications agre...

Page 3: ...nsing lines 4 2 Electronic connection By authorized and qualified specialized person nel only The electrical connection of the instrument shall be performed according to relevant VDE and local electri...

Page 4: ...re menu select Quit and press Example Setting the switch points In normal mode press the button to enter the menu Menu level Switch points appears in the display to change the switch points press the...

Page 5: ...measuring cell corresponds to that of a capillary throttle You can set the response time to pressure jumps in the range 0 0 to 100 s But with maximum damping it will take more than 2 minutes for the r...

Page 6: ...EAR is active the other menu items are cancelled Square rooted Here the input signal is square rooted before being sent to the display and the out put This is necessary e g for flow measurements with...

Page 7: ...nt unwanted colour changes Here the colour change can be delayed in the range between 0 100s The following applies for instruments with firmware 2 16 and later Here the colour change can be delayed in...

Page 8: ...wn stream components Check the leak tightness of the pressure con nection lines Check the electrical connections The exact test cycles shall be adapted to the oper ating and ambient conditions If vari...

Page 9: ...g voltage Ub 20 32 V DC AC Connection type Three wire Output signal 0 20 mA 4 20 mA 0 10 V DC Apparent ohmic resistance Ub 26V RL Ub 4V 0 02A Ub 26V RL 1100 RL 2 k Power consumption approx 2 W VA Meas...

Page 10: ...nless otherwise specified 90 75 25 5 M12 plug connection Wall mounting plate optional Rear view without wall mounting plate 70 5 61 5 4x boreholes for Tapping screws 3 5 Hose screw connection for 6 or...

Page 11: ...DC AC 20 32 V DC AC N Measuring unit Selectable pressure units W Measured value display contact elements 4 digit colour change LCD 2 relay contacts C 4 digit colour change LCD 2 semiconductor switch...

Page 12: ...without notice Changements techniques sous r serve Fischer Mess und Regeltechnik GmbH Bielefelder Str 37a D 32107 Bad Salzuflen Tel 49 5222 9740 Fax 49 5222 7170 eMail info fischermesstechnik de www...

Reviews: