25
Table 11 - Camera LED indication and Description
Camera status
Camera LED indication
Description
Operational
Purple blink
The camera is operational
Operational cooling
White blink
The sensor is being cooled
Operational throttling
Double purple blink
The camera cannot reach the target
temperature
Safe
Red blink
An error has been detected. The detector
is turned off. The camera needs to be
rebooted.
Note: turned off led does not necessary mean that there is an issue with the camera. The camera can
be configured to switch off the led automatically once the boot is completed.
Power supply
C-BLUE One can be powered either using Power-over-CoaXPress
®
(PoCXP), with a compliant
CoaXPress
®
frame grabber, or using a dedicated power input, supplied in the delivery package.
Note: For optimal thermal performances, it is recommended to use the external power supply via
the LEMO® connector.
Via PoCXP
Refer to 8.3.6 for a complete description of the Power-over-CoaXPress
®
function.
External power supply
The provided power supply provides a stable 12 V DC, with at least 5 A of current available (60 W) that
allows to properly power C-BLUE One when cooling is enabled.
The mating connector is a LEMO
®
FGG.0B.304.CLAD52Z
C-BLUE One back view.
Always use the power supply provided by First Light Imaging. Using another power
supply can damage the camera.