9.2.1 Measurement Settings Menu
Setting
1.1
Unit
Application
Probe Length
Adjustment L.V.
Measurement Settings (1) includes 6 parameters for setting, as introduced below:
Unit Type (1.1): It is to set the unit of measurement value shown on the LCM screen.
Probe Length (1.2): It is to set the length of sensing probe.
Application (1.3): It is to set the application model and dielectric constant of the media to be
measured.
Adjustment L.V. (1.4): It is to set the upper/lower limit corresponding to 4-20mA and the
percentage.
Filter Level (1.5): It is to set the output filter level.
Current Set (1.6): It is to set the current output direction (4~20mA/20~4mA).
Setting
1.1
Unit Type
Application
Probe Length
Adjustment L.V.
Unit Type
m
mm
cm
1.1.1
in.
In the Unit (1.1) menu, user may adjust the unit of measurement value shown on the LCM
display, with 5 available units, m/cm/mm/in/ft.
Measurement Settings
1.2
Unit Type
Application
Probe Length
Material Level Adjustment
Probe Length
1.2.2
YES
NO
Auto detect?
Probe Length
0 6 4 0 0
+0.0000 ... +30000
mm
Probe Length (1.2) is applied in the bottom tracking probe mode, which can be filled after
the probe is changed, or detected automatically by placing the instrument in an empty
tank.
Note: The material height shown on the instrument is the probe length deducted with the
distance from the material level to the flange reference plane.