3
3-12
Color Calibration
N
OTE
:
ColorWise Pro Tools for Windows and Mac OS computers are fundamentally
the same; differences are noted in this chapter. The screens and dialog boxes you see
illustrated here are the Windows version.
You will need the Color Reflection Reference strip included with the
spectrophotometer. Remove the strip from its protective envelope, handling it only by
the edges. Keep the strip free of dust and smudges, and always store it in the protective
envelope.
T
O
CALIBRATE
THE
X-R
ITE
DTP41
1.
Connect the spectrophotometer to the computer and turn on power (see page 3-8).
2.
Launch ColorWise Pro Tools and connect to the Fiery 3850C.
For instructions on configuring the connection to the Fiery 3850C, see Getting Started.
N
OTE
:
If the connection to the Fiery 3850C from Command WorkStation that also
allows for access to ColorWise Pro Tools is not established, you can directly launch
ColorWise Pro Tools independently of Command WorkStation.
3.
Click Calibrator.