background image

Infrared 
Refrigerant
Leak Detector
Model: SRL2

OPERATOR’S MANUAL

Description

The  SRL2  uses  infrared  optics  to  create  a

portable  refrigerant  leak  detector  technology  that
has  a  superior  combination  of  sensitivity,  speed,
sensor life, battery life, portability, and ease of use.  

Gas is pumped through the tip of the wand into

the  sensor  within  the  SRL2  body.  The  sensor
detects changes in concentration, not absolute con-
centration,  making  it  easy  to  detect  leaks  even  in
areas with refrigerant in the air. 

The SRL2 detects leaks as small as 0.1 oz/year,

which exceeds the toughest SAE J1627  standards.
The SRL2 comes with wall and car charger for its
ultra-compact Lithium-ion battery which powers the
SRL2  for  8hrs  of  continuous  operation  before  a
recharge is needed. That’s long enough to last your
entire workday.

Unlike  many  leak  detectors,  oil  vapor  does  not

trigger the SRL2.

The  SRL2  has  vari-

able  sensitivity  set-
tings,  keeping  ‘nui-
sance  tripping’  to  a
minimum.  A  quick  and
automatic 30 second self-calibration upon power-up
ensures optimal performance. A built in replaceable
filter blocks moisture and harmful particulates. The
SRL2 also has a mute button and a peak hold func-
tion.

Quick Tips

1.

Keep the wand tip moving past suspected leak
locations.

2.

Once a leak is detected, sweep the wand back
over to pinpoint.

3.

To pinpoint larger leaks adjust sensitivity.

Filter Assessment & Replacement

The filter blocks moisture and other contaminants

from the sensor. When it gets wet, it constricts flow
of air and it will have to be replaced. Unscrew the
sensor  tip  and  replace  the  white  filter  so  that  the
rounded end is closest to the tip of the wand. Use
only the Fieldpiece supplied filter. 

You  can  order  from  a  distributor  extra  bags  of

replacement parts (model RFL2) are available from
Fieldpiece. Model RFL2 contains 10 filters and 5 O-
rings. 

Comparing Leak Detectors

Fieldpiece has two leak detectors. 

The SRL2 uses an infrared (IR) sensor. It detects
a CHANGE in concentration, so the wand 
must keep moving. The main advantages of the
Fieldpiece IR sensor are that its sensitivity will 
remain the same over the life of the instrument, 
the sensor will last the lifetime of the instru
ment, it will not trigger on humidity or on oil, 
and it has nearly the same sensitivity to most 
refrigerants.The main disadvantage of infrared 
sensors is their sensitivity to mechanical 
disturbances.

The SRL8 uses a heated diode sensor. 

The heated diode'sadvantage is that it detects
absolute concentration levels so it can be held on
a leak and will continue to beep. It is also initially
very sensitive. Its disadvantages are that the 
sensor becomes less sensitive over life, eventually
needing replacement and it is more sensitive to
some refrigerants than others.

T2

T2

T2

T2

T2

T2

T2

T2

T2

T2

T2

T2

T2

T2

T2

T2

PEAK

MUTE

ON/OFF

ON/OFF

L/M/H

SENSITIVITY

H

M

L

PEAK

LOW-BATT

PRESS FOR

1 SECOND

SRL2

Infrared

Refrigerant

Leak Detector

BATTERY

CHECK

SENSITIVITY

H

M

L

PEAK

LOW-BATT

 

 

 

Operation

ON/OFF Protection

To  turn  on/off  the  SRL2  press  and  hold  the

ON/OFF  button  for  one  second.  This  slight  delay
protects  against  inadvertent  pressing.  If  you  forget
to turn it off, it will automatically go off in 10 minutes.

LED Bar Graph Display

The  eight  segment  LED  display  indicates  the

degree  of  change in  refrigerant  concentration.  As
the concentration of refrigerant in the air increases,
so does the number of lit bars on the display.

L/M/H Button (Sensitivity)

Set the sensitivity level by pressing the L/M/H but-

ton.  Low(L),  medium(M),  or  high(H)  sensitivity  will
be indicated by their respective LED. 

The higher the concentration of refrigerant in the

ambient air, the lower the sensitivity setting should
be to minimize false trips.

Mute Button and Battery Check

Pressing  the  MUTE  button  toggles  the  sound  of

the SRL2 on and off. In addition, holding the MUTE
button down for one second will display the remain-
ing percent of battery charge on the LED bar graph. 

Peak Button

The  PEAK  function  holds  the  highest  change  in

concentration  achieved  while  continuing  to  detect
leaks. Press PEAK to toggle this function on and off.
Turning  the  PEAK  function  off  will  clear  the  peak
change. The PEAK LED will light when PEAK func-
tion is on. 

Turbo Mode

The TURBO function is an additional 4th sensitiv-

ity that pushes the SRL2 to it highest possible sen-
sitivity.  This  feature  is  activated  by  pressing  the
PEAK button four times in succession, and can only
be activated while on the high(H) sensitivity level. 

When in the TURBO mode the first green LED on

SRL2 will strobe, and the audio detection sound will
be a solid tone. 

To exit the TURBO mode, press the PEAK button

again  four  times  in  succession,  or  switch  from
high(H) to one of the other sensitivities levels.

Beware:  While  in  TURBO  mode  the  SRL2  is

extremely  sensitive  and  will  more  easily  trigger  on
abrupt  or  violent  motions  as  well  as  very  small
changes  in  refrigerant  levels.    Handle  with  care
when in TURBO mode.

Leak Detection Procedure

Keep the SRL2 away from any areas of potential

refrigerant leakage until the warm-up and calibration
period  is  over.  The  warm-up  and  calibration

sequence  lasts  about  30  seconds  after  ON/OFF  is
pressed.  Allow  the  SRL2  to  fully  warm-up  before
using  to  avoid  false  triggering.  On  startup,  meter
defaults to high sensitivity.

The most likely points for refrigerant leaks are at

the soldered joints in refrigerant lines and changes
in cross section or direction of these lines.  

The  SRL2  detects  changes in  concentration  of

refrigerant, not the absolute concentration of refrig-
erant. This lets the user easily detect leaks in places
that may have refrigerant in the air, such as a refrig-
erator leaking into an enclosed space. Because the
SRL2  detects  change,  there  is  a  specific  “double-
pass” method used to find leaks. 

1. The tip of the wand should be fairly close to the

line. You may need to be within 1/4” (0.63cm) of
a small leak to detect it. In this case, using a
second hand to guide the tip along refrigerant
lines may be helpful.

2. Keep the tip moving along refrigerant lines at a

rate of 1-3 inches per second.

3. Once the SRL2 indicates a change in concentra-

tion, note the spot on the line and keep the tip
moving past the potential leak to refresh the air
space within the SRL2 with clean air. (Note: If
the SRL2 wand moves through a very high con-
centration of refrigerant, you may need to use in

fresh air for 4 seconds before moving to step 4
in order to clear the refrigerant within the SRL2.)

4. Return the tip back to, and past, the spot first

indicated. When the SRL2 indicates a second
change, note the spot on the line. The source of
the leak will be near the midpoint between the
two noted spots of indication. 

5. Very high concentrations of refrigerant can

cause an overload which can take a few sec-
onds to clear. 

SRL2 testing an A-coil in an evaporator, 1/4” from the line.

1/4”

(0.63

cm)

WARNING

DO NOT use the SRL2 without the proper fil-

ter correctly installed. 

!

!

Reviews: