1
www.fiduspower.com
MVPS600
Series
Model Number
(1)
Output Power
(2)
Output Voltage
Output Current (Natural Convection)
U Channel
Slotted Cover
Solid Cover
MVPS600
-
1
0
12
180
-
300W
12V
25A
17.5A
15A
MVPS600
-
1
0
15
225
-
375W
15V
25A
17.5A
15A
MVPS600
-
1
0
24
360
-
600W
24V
25A
17.5A
15A
MVPS600
-
1
0
30
360
-
600W
30V
20A
14A
12A
MVPS600
-
1
0
48
360
-
600W
48V
12.5A
8.75A
7.5A
MVPS600
-
1
0
58
360
-
600W
58V
10.34A
7.25A
6.2A
Models & Ratings
Dimensions:
5 x 8.5 x
1.61” (127
x 216 x 40.9mm)
Parameter
Minimum
Typical
Maximum
Units
Notes & Conditions
AC Input range
85
264
VAC
Derate from 100% at 120VAC to 70% at 85VAC
Operating temperature
-
40
70
°C
See derating curve p3.
Efficiency
120VAC 88% and 240VAC 93% typical
Dimensions
5 x 8.5 x
1.61” (127
x 216 x 40.9mm)
EMC
EN55011 Level B conducted and radiated with king core K5B RC 25x12x15
-
M on input. EN61000
-
3 and EN61000
-
4, harmonics,
flicker, Surge, EFT, ESD, conducted and radiated immunity to IEC 60601
-
1
-
2 4th ed
Safety
EN60601
-
1, IEC60601
-
1 (ed3), ANSI/AMMI ES 60601
-
1, CSA C22.2 No 60601
-
1
Key specifications
•
5 x
8.5”
Footprint,
1.61”
for 1U installation
•
Latest medical approvals IEC60601
-
1 (3.1) IEC60601
-
1
-
2 (4th ed), BF rated
•
Up to 3 units load share for higher powers
•
Peak of 1000W for 500mS
•
EN55011 Level B conducted & radiated
•
5 Year warranty
600 Watts
The MVPS600 series of low profile, medically approved AC
-
DC power modules offer 300
-
600W in a
5”
x
8.5”
package conforming to the latest IEC60601
-
1 3.1 and IEC60601
-
1
-
2 4th ed standards. The
units are fully featured including 5V 1.5A auxiliary output and low standby power of 3W. The units are
body floating (BF) rated for applied part applications. They have a wide operating temperature of –
40
to 70°C, in a range of voltages from 12V to 58V. All units come with a FiDUS 5 year warranty.