10.4.5 Relay output
If optional relay output is equipped: -
•
Relay 1 output synchronizes with buzzer.
•
Relay 2 output synchronizes with Lo.
•
Relay 3 output synchronizes with OK.
•
Relay 4 output synchronizes with Hi.
Note: -
Set internal function F15 to enable/disable buzzer output.
10.5 Memory Accumulation
Below memory accumulation modes are available for each Uart port: -
a.
Memory accumulation disabled.
b.
Manual accumulation.
c.
Automatic accumulation.
Notes: -
•
During counting operation, both net weight and piece value will be accumulated.
•
Date stored will not be erased when powered off.
•
Data stored will be erased when operator code is changed.
10.5.1 Memory accumulation settings
To enable/disable memory accumulation, an Uart should be set to one of the following modes: - Auto 1; Auto 2;
Auto 3; Manual.
•
To enable memory accumulation, set Auto Accumulation to On in internal function number F16.
•
To disable memory accumulation, set Auto Accumulation to Off in internal function number F16.
Refer to step
4
of
paragraph
8.3
for setting details.
10.5.2 Manual accumulation
Press
[Print/M+]
to accumulate the current value to memory.
10.5.3 Automatic accumulation
13
Automatically memory accumulation of an Uart is activated automatically when
Auto Accumulation
is set to
On
in Auto 1, Auto 2, Auto 3. Under the automatic accumulation mode, corresponding results will be
accumulated automatically.
Unstable results or results which are less than 20d (or 20d
1
for dual weighing range/interval mode) will not be
automatically accumulated.
10.5.4 When data is accumulated to memory
When a result is accumulated to memory: -
•
Instrument displays “
n____x
” on LCD3.
•
M+ Indicator
appears to indicate that memory contains stored data. “x” means the total number of
transactions accumulated to memory.
•
This instrument returns to normal display status after 2 seconds.
13
Auto 1 ~ 3 do not support PA.