- 8 -
Splice Programs
Splice Menu, Maintenance, Stabilize Electrodes, Setting, User Instruction, Arc Calibration
Atmospheric conditions such as temperature, humidity, and pressure are constantly changing, which
creates variability in the arc temperature. This splicer is equipped with temperature and pressure
sensors that are used in a constant feedback monitoring control system to regulate the arc power
at a constant level. However, changes in arc power due to electrode wear and glass adhesion cannot
be automatically corrected. Also, the center position of arc discharge sometimes shifts to the left or
to the right. In this case, the fiber splicing position has to be shifted in relation to the arc discharge
center. It is necessary to perform an arc power calibration to eliminate these problems.
Note : Performing [Arc calibration] function changes the arc power “Factor” value. The factor value is
used in the algorithm program for all splicing. The arc power value will not change in the splice modes.
* Standard Factor value is changed within 11 ± 2, It shows “Complete” word.
[Arc Calibration]
In the event of sudden change in environmental conditions or after cleaning electrodes, the arc power
sometimes becomes unstable, resulting in higher splice loss. This is especially a concern when the
splicer is moved from lower altitudes to higher , it takes time for the arc power to stabilize. In this case,
stabilizing electrodes will expedite the process to normalize the arc power. If many tests are required
to get the “Test ok” message appearing the [Arc calibration], use this function as well.
[Stabilize Electrodes]