Section 2: Installation
2.6 Cable Installation
0101-8242-0, Rev. C
2-13
CV-6SLX User Manual
2.6.5 Making I/O Connections to the Vacuum System
NOTE
Before performing this portion of the installation, complete the mandatory
baseline test described in Section III of document 0101-8241-1, the CV-6SLX
Quick Start Guide.
The installer must make connections (1) between the vacuum system and the HVPS I/O cable
(PN 0620-6682-0) and (2) between the vacuum system and the FPS I/O cable (PN 0620-6672-
2). Control inputs and status outputs include both digital and analog signals. Digital inputs
require a simple contact closure and carry 24 V dc at approximately 20 mA. All digital outputs
are +24 V dc sourcing a maximum of 50 mA. For detailed information about I/Os exchanged
via the HVPS I/O cable, see Figure 2-15 and Table 2-1. For details about I/Os exchanged via
the FPS I/O cable, see Figure 2-16 and Table 2-2.
When making I/O connections to the HVPS I/O cable, use either the optional 25-pin interface
module available from Temescal (PN 6149-2293-637) or a comparable interface module.
When making I/O connections to the FPS I/O cable, use either the optional 15-pin interface
module available from Temescal (PN 6149-2293-624) or a comparable interface module.