35X-NM Non-Magnetic Cot

© Ferno-Washington, Inc. / 234-3494-01 / July 2018

3

Patient Restraints

SAFETY INFORMATION

The cam buckles on the three restraints provided with the cot are made 

from plastic and are suitable for use in an MR environment but are not 

intended to be placed inside the bore of the MR scanner. 

NEVER

 use 

metal-buckle restraints in an MR environment.
Primary restraints (rated for use in motor vehicles; not provided) must 

be used when the cot is in an ambulance. 

NEVER

 use the cam-buckle 

restraints to secure a patient for transport in an ambulance. To order 

restraints, contact Ferno. See “Ferno Customer Relations” on page 8.

GENERAL GUIDELINES FOR USE

● 

Always use a complete set of three patient restraints to secure the 

patient on the cot. Adjust restraints to safely secure the patient 

without causing discomfort or impairing circulation.

● 

Keep restraints fastened when not in use to prevent them from 

interfering with cot operation.

ATTACHING CAM-BUCKLE (PLASTIC) RESTRAINTS

1.  Select attachment points for the cam-buckle restraints, ensuring 

solid anchorage for the straps and proper restraint of the patient 

(Figure 1). Verify that restraints will not interfere with use of the 

cot or accessories.

2.  Unbuckle a restraint.
3.  Slide the looped end of a strap around the selected attachment 

point and thread the buckle or loose end of the strap through the 

loop (Figure 2). Pull the loop tight around the  attachment point.

4.  Repeat Steps 1-3 with the remaining five straps.
5.  Fasten the restraints.

USING CAM-BUCKLE (PLASTIC) RESTRAINTS

● 

To fasten a cam-buckle restraint: 

Open the cam lock and thread 

the strap through the opening (Figure 3), then close the cam lock 

to hold the restraint in place (Figure 4).

● 

To unfasten a cam-buckle restraint: 

Open the cam lock and 

pull the restraint backward through the open lock.

USING METAL-BUCKLE RESTRAINTS

Refer to the users’ manual for the restraints and/or the Model 35X cot for 

proper attachment points, use and care.

Figure 1 - Example of Restraint Placement

Figure 2 - Attaching a Cam-Buckle Restraint

 

NOTICE

Different styles of patient restraints are required in different 

situations.
When the cot will be used in an MR environment, use only 

the plastic, cam-buckle restraints provided. Metal-buckle or 

other styles of restraint may not be suitable for use in the MR 

environment.
When the cot will be used in an ambulance, use only primary 

restraints rated for use in a vehicle. Cam-buckle restraints are not 

suitable for restraining a patient in a vehicle.

 

WARNING

An unrestrained patient can fall off the cot and be injured. 

Use restraints to secure the patient on the cot.

Figures 3 and 4 - Fastening a Cam-Buckle Restraint

4

3

Summary of Contents for 35X-NM PROFlexx

Page 1: ...equirements of the Star of Life certification via Federal Ambulance Specification KKK A 1822 For information contact Federal Supply Services Specifications Section Suite 8100 470 E L Enfant Plaza SW Washington DC 20407 Purpose of This Manual This manual Ferno document 234 3494 is a supplement to the 35X PROFlexx Series Users Manual The specific features of the 35X NM cot are described in this docu...

Page 2: ... product that is Magnetic Resonance Conditional called MR Conditional in this manual and on product labeling has been shown to pose no hazards in a specified MR environment under listed conditions of use LISTED CONDITIONS OF USE The Model 35X NM cot has undergone testing and has been determined to be MR Conditional In general the cot is intended for use under the conditions listed here Static magn...

Page 3: ... 3 Slide the looped end of a strap around the selected attachment point and thread the buckle or loose end of the strap through the loop Figure 2 Pull the loop tight around the attachment point 4 Repeat Steps 1 3 with the remaining five straps 5 Fasten the restraints USING CAM BUCKLE PLASTIC RESTRAINTS To fasten a cam buckle restraint Open the cam lock and thread the strap through the opening Figu...

Page 4: ... locked into the new position After Use Buckle and or adjust the torso restraint Sidearms Purpose Sidearms provide patient security and comfort Keep the sidearms raised except when transferring a patient to or from the cot The sidearms feature an intermediate stop that allows a patient s arm to be placed on the sidearm for treatment starting of intravenous IV fluids etc Use To lower pull the relea...

Page 5: ...Cot Before transferring the patient to the cot confirm the style of restraint on the cot is correct for the cot s use See Patient Restraints on page 3 Use only the plastic cam buckle restraints provided when the cot will be used in an MR environment Use only primary restraints such as metal buckle restraints when the cot will be used in an ambulance Figure 7 Wheel Lock NOTICE Different styles of p...

Page 6: ...the cam buckle restraint buckles and webbing following the disinfectant manufacturer s directions for application method and contact time Cleaning Place the cam buckle restraints in a mesh bag not supplied and machine wash in warm water with a mild detergent Hang the restraints to dry Do not place the restraints in a dryer METAL BUCKLE OR OTHER RESTRAINTS Disinfect and clean only as directed in th...

Page 7: ... tubing Do not lubricate Wheels transport and loading Do not lubricate Center leg crosstube Do not lubricate Backrest hinge Do not lubricate Telescoping side handles Do not lubricate Telescoping legs Do not lubricate Shock frame hinge Do not lubricate Control handle mechanisms Do not lubricate Applying Lubricant Use a small amount of lubricant one or two drops Lubricate identical points on each si...

Page 8: ...itions of the limited warranty and the limitations of liability and disclaimers are also available upon request by calling Ferno at 1 800 733 3766 or 1 937 382 1451 Repairs U S A and Canada EMSAR istheonlyorganizationauthorizedbyFerno to manage service and repair Ferno products Telephone Toll free 1 800 73 EMSAR Telephone Worldwide 1 937 383 1052 Fax Worldwide 1 937 383 1051 Internet www EMSAR com...

Reviews: