Users’ Manual
May 2018
Pub. No. 234-3398-02
Read this Manual and Retain
for Future Reference
Model 30-NM
Non-Magnetic Cot
Page 1: ...Users Manual May 2018 Pub No 234 3398 02 Read this Manual and Retain for Future Reference Model 30 NM Non Magnetic Cot...
Page 2: ...are covered by a limited warranty which is printed on all Ferno invoices The complete terms and conditions of the limited warranty and the limitations of liability and disclaimers are also available...
Page 3: ...Handle______________________________________ 12 4 6 Backrest__________________________________________ 12 4 7 Wheel Lock_______________________________________ 12 5 Using the Cot______________________...
Page 4: ...o secure the patient on the cot Rolling the cot sideways can cause it to tip and injure the patient or operators Roll the cot only with the loading end or control end first Improper maintenance can ca...
Page 5: ...ide the MR system e g inside the bore of the scanner FERNO COMPLIANCE Ferno adheres to the ASTM International classification and definitions found in the following document The American Society for Te...
Page 6: ...and equipment into and out of the ambulance Use additional help as needed to lift the weight of the cot patient and equipment See Using Additional Help on page 7 Operator Focus WARNING Untrained oper...
Page 7: ...ce Refer to the chart for suggested placement for operators and helpers WARNING Helpers can cause injury Maintain control of the cot operate the controls and direct all helpers Helpers can be injured...
Page 8: ...mpartment must have a level floor large enough for the folded cot and a cot fastener installed not supplied The cot has been shown to pose no hazards in a specified MR environment under listed conditi...
Page 9: ...s All Ferno mattresses for emergency medical service use are fully sealed Specifications Imperial Metric Load Capacity 400 lb 181 kg Load Capacity UK 28 5 st Weight 54 lb 24 5 kg Cot Construction Tubu...
Page 10: ...nly in an MR environment Use metal buckle restraints only in an ambulance 4 1 Patient Restraints Use a complete set of three patient restraints to secure the patient on the cot Adjust restraints to sa...
Page 11: ...ositions on page 14 for complete instructions Control Handles 2 4 3 Swing Down Sidearms Sidearms help provide patient security and comfort Keep sidearms raised except when transferring a patient onto...
Page 12: ...k A wheel lock is fitted into all four transport wheels Figure 7 Wheel locks help keep the cot stationary during patient transfer and certain medical procedures When engaged the wheel lock stops wheel...
Page 13: ...have a Ferno fastening system installed 5 2 General Guidelines for Use Medical advice is beyond the parameters of this manual It is the users responsibility to ensure safe practices for the patient a...
Page 14: ...engages if it has not already done so 6 Both Operators Set the cot onto the floor or ground Do not release your grasp until your are sure the undercarriage is locked at the new position CHANGING POSIT...
Page 15: ...perators and helpers should grasp the cot with both hands ROLLING THE COT 1 Securely fasten the patient restraints around the patient 2 Both Operators Place the cot in a rolling position 3 Both Operat...
Page 16: ...minimum of two trained operators who are communicating working together and maintaining control of the cot at all times Note The control handles are not used during this process 1 BothOperators Prepa...
Page 17: ...quid 3 Webbing Add a disinfectant cleaner to warm water following the disinfectant manufacturer s instructions Immerse webbing in the solution for the time directed by the disinfectant manufacturer Do...
Page 18: ...ces of the cot with warm water and a mild detergent 3 Rinse with warm clear water Dry the cot with a towel or allow it to air dry 6 6 Waxing the Cot While it is not necessary to wax this cot waxing th...
Page 19: ...ead Handle WRL 191S or E Z 1 2 Upper ratchet bar top and bottom WRL 191S or E Z 1 3 Sidearm lock pins WRL 191S or E Z 1 4 Telescoping frame WRL 191S or E Z 1 5 Backrest hinge WRL 191S or E Z 1 6 Slidi...
Page 20: ...nal repair contact your Ferno distributor Your distributor is the only agent authorized by Ferno to manage service and repair Ferno products Parts and Service PARTS LIST No Part Description Qty 1 190...
Page 21: ...ems to the cot can cause injury Use only Ferno approved items on the cot Description Part Number Mattress Model 360 flat orange 031 3691 Mattress Model 360 flat burgundy 031 3692 Mattress Model 460 bo...
Page 22: ...Ferno Washington Inc 234 3298 02 May 2018 22 Training Record Date Printed Name Signature Training Method Trainer Initials Read Manual Video Online Hands On TRAINING RECORD...
Page 23: ...Ferno Washington Inc 234 3298 02 May 2018 23 Date Maintenance Performed By Maintenance Record MAINTENANCE RECORD...
Page 24: ...Ferno Washington Inc 234 3298 02 May 2018...