Advanced Display (AD.DI) -

 Segments linearization. The 

instrument has 10 linearization segments (11 points). Each 
linearization points is formed of a signal input value (“INPX”) 
and an associated display value (“DISPX”). Enter the values 
for each point from the Points menu (“PNTS”). It is required 
that the input signal values “INPX” must increase for each 
new point. When all points are entered, select “Act” value 
to “On” in the activation menu. A check is automatically 
performed when entering the points and when activating the 
points. In case of error verify the data entered on the points. 
Leave the menu pressing “LE” button. Reset option “REST” 
deletes all existing points.

Password (PASS) - 

Select a number to act as password. This 

password will be requested when entering the Configuration 

Menu. To deactivate the password select “Off”.

Factory Settings (FACT) - 

Factory  default  configuration. 

Select “yES” to activate the factory default configuration.

Version (VER) - 

Firmware version installed.

Light (LIGH) - 

Brightness.  Select  between  5  predefined 

levels of brightness.

Menu options OPT1 and OPT2 give access to the configura

-

tion menus of the options installed at slots Opt1, Opt2 and 
Opt3. This menu depends on the installed option. If there is 
no option installed the instrument shows “

NONE

”. Control 

modules R1 are controlled from the standard alarm menu 
(see section 1.12.7).

Peak & Hold (P.HLD) 

- Peak & Hold function detects sharp 

drops in the displayed value, and holds the display if a display 
drop is detected. To disable the hold function for a moment, 

press any of the front buttons or wait for a predefined time :

   Time 0 

 

hold disabled (Off)

   Time  1 a 3999    

seconds, before disabling the hold

   Time 4000 

 

infinite hold

The counting of seconds is started each time there is an in-
crease in the display value. Alarms will follow the input signal 
while hold is active.

1.10.7 Tools

1.10.8 Menu OptX - Options

When the instrument detects that the displayed value does 

not correspond to the expected value, the display will flash 

and alternate with a message.

h.udr

”  Hardware underrange. The input signal is negative.

h.oVr

”  Hardware overrange. The input signal is higher 

than excitation voltage (+5V).

d.udr

”  Display underrange. The instrument is displaying 

the minimum value (-1999) and can not display below.

d.oVr

”  Display overrange. The instrument is displaying the 

maximum value (9999) and can not display above.

hoLd

”  The instrument is showing the value present when 

the hold function was activated. Hold function is active.

Min

” 

The instrument displays the minimum displayed 

value in memory. The minimum visualization is active.

MAX

”  The instrument displays the maximum displayed 

value in memory. The maximum visualization is active.

P.hLd

”  The instrument displays the peak value. Function 

Peak&Hold is active.

Err.0

 

Values introduced on the “ScAL” menu are not valid. 

Default values are activated. The slope defined by the two 

points is almost vertical (higher than 5000):

Err.1

 

Password incorrect.

Err.2

 

The instrument has detected an installed option but 

was unable to communicate.

Err.3

 

Related to the edition of a linearization point. Re-

turns to the point for further edition. The value of “InPX” of 
the edited point is lower than “InPX” of the previous point. 

Err.4

 

Related to the activation (“Act”) of the lineariza-

tion segments (“Ad.dI”). Activation is not allowed. The value 
“InPX” of one of the points is higher than “InPX” of the next 
point.

Err.5

 

Related to the activation (“Act”) of the linearization 

segments (“Ad.dI”). Activation is not allowed. The slope de-

fined by one of the segments is almost vertical.

Err.8

 

Excitation voltage overloaded. Excessive current 

is being demanded to the excitation voltage. Potentiometer 
nominal value is lower than 150 Ohms.

E.101

 

Option is installed but the type can not be recog-

nized.

(dhI-dLo)  [Counts]
(IhI-ILo)   [mA or Vdc]

< 5000

Slope condition for Err.0 and Err5

1.11 Messages and errors

Summary of Contents for S Series

Page 1: ...earization maximum and minimum display memory scalable reading filters steps signal autocorrection measure function peak hold alarms with double setpoints 5 levels of brightness Universal AC and DC po...

Page 2: ...dard IP54 front protection with optional upgrade to IP65 protection Connections via plug in screw terminals and configuration via three front push buttons For application on industrial environments S4...

Page 3: ...wer line a protection fuse acting as disconnection element easily accessible to the operator and identified as a protection device Power H fuse 250mA time lag Power L fuse 400mA time lag 6 7 8 9 0 1 2...

Page 4: ...ec 0 to 99 signal Power Power H 85 to 265 Vac dc Power L 11 to 60 Vdc and 24 48Vac Consumption 4W Isolation 3500Veff for power H 2000Veff for power L all levels tested for 60 seconds Configuration 3 f...

Page 5: ...ue outputs The Information Menu is for information only and it does not accept changes on the displayed information To enter the In formation Menu press the UP button It is not affected by the PASSWOR...

Page 6: ...ut High Display High Decimal Point Alarm1 Setpoint Hysteresis Delay Setpoint2 Counts Seconds Second setpoint Auto correction Auto High Auto Low Alarms Signal Low Display Low Signal High Display High R...

Page 7: ...zation Password Version Points Enter signal values Input and associated display val ues Display for each linearization point Then activate Act to On In case of error verify the points data Select pass...

Page 8: ...alarm will be activated between 1000 and 1500 and the alarm will be deactivated when display is 1000 and 1500 Setpoint2 is affected on the same way as the setpoint with hysteresis and delays Function...

Page 9: ...al while hold is active 1 10 7 Tools 1 10 8 Menu OptX Options When the instrument detects that the displayed value does not correspond to the expected value the display will flash and alternate with a...

Page 10: ...he configured input signal range 4 20mA 0 10Vdc and the values for input low ILo display low dLo input high IhI and display high dhI Maximum MAX Value of the maximum display Minimum MIn Value of the m...

Page 11: ...to the 61010 1 CE security regulation and is designed for applications on industrial environ ments See the CE Declaration of Conformity further in this document for information on the category of meas...

Page 12: ...AO modules can be installed in one S Series panel meter For more information see document 2657_S40_OPTIONAL_ MODULES at www fema es Module with 1 relay Up to a maximum of two R1 modules can be instal...

Page 13: ...contacts Common NO NC Maximum current 6A resistive load each relay Voltage 250 Vac continuously Installable at slot Opt 1 R2 fills OPT1 R4 fills OPT1 and OPT2 Terminal Plug in screw terminals pitch 3...

Page 14: ...User s Manual S40 R FEMA ELECTR NICA Page 14 T H I S P A G E B L A N K...

Page 15: ...User s Manual S40 R FEMA ELECTR NICA Page 15 T H I S P A G E B L A N K...