Operation

 

5-3. 

Function of judging whether the detection tube is clean

 

This function judges whether the detection tube is clean when the washing process ends after manual/auto operation. 

 

 When the washed detection tube is not clean

 

When it is judged the detection tube is not clean, "Warning: cell too dirty" is displayed on screen. 

When this message is displayed, manually wash the detection tube (see “8-2-1”), shut off the primary power supply and clean the sensor 

(with brush, see”8-4”).    After that, the primary power supply again and perform manual running using mineral water (see “8-1-1”).   

Since stain level of sample water detection tube is judged after manual operation is finished, this message disappears when the detection 

tube is clean.    If the message is still displayed, repeat the above steps. (clean the sensor (with brush, see”8-4”), manual operation using 

mineral water (see “8-1-1”) and judging the stain level of sample water detection tube).   

 

 

9CTPKPI㨍GNNVQQ&KTV[
ZZZZZZZZZZZZ

 

 When th-e inside of the detection tube is too dirty to be measured

 

"Warning [112]" is displayed on screen when the inside of the detection tube is too dirty to be measured. 

Clean the sensor (with brush, see”8-4”) to maintain the performance of the detector.    It is necessary to temporarily cancel the Warning state 

to enable sensor cleaning (with brush, see”8-4”).    Press “Enter.”    Then display changes to “Warning: cell too Dirty.”    Shut off the primary 

power supply, and clean the sensor (with brush, see”8-4”).    After that, supply the primary power supply again and perform manual 

operation using mineral water. (see “8-1-1”)    Since stain level of sample water detection tube is judged after manual operation is finished, 

this message disappears when the detection tube is clean.    If the message is still displayed, repeat the above steps. (cleaning the sensor 

(with brush, see “8-4”), manual operation using mineral water (see”8-1-1”) and judging the stain level of sample water detection tube).   

 

9CTPK㨘I=?
#FJGUKQPQHFKTV

 

%#76+10

 

Ɣ

 When warning message still remains on screen after the sensor cleaning (with brush), the fresh water may be 

contaminated.    Do not use industrial water or circulating reclaimed water in the fresh water line since such water may 

be cloudy or contaminated. 

Ɣ

 

This 

oil concentration detector is equipped with a “Zero-point compensation” function.    However never use this function 

when “Warning: cell too dirty” or ”Warning [112]” (to alarm dirtiness of sample water detection tube) is displayed on 

screen.    If used, a detector trouble may occur.

 

 

Summary of Contents for FOCAS-2000

Page 1: ...Concentration Detector Instruction Manual Read this instruction manual thoroughly before use to operate the product safely by following precautions Be sure to keep it in an easy to access place so as...

Page 2: ...Test operation method by theAutomatic operation function page 13 5 3 Function of judging whether the detection tube is clean page 14 6 Setting 6 1 Tabulated list of menu items page 15 6 2 Time correc...

Page 3: ...comesinsufficient this machine tells it on the LCD display When the free space of the operational log memory becomesinsufficient the machine may be disabled to run To run the machine you must replace...

Page 4: ...t method Optical continuous measurement Operation method Automatic operation using sample water pressure Response time 5sec or less comply with MEPC107 49 Power supply power consumption AC90 240 V 50...

Page 5: ...aximum 354 AL 1 POWER 15ppm Bilge Alarm Esc Enter AL 2 A MODEL FOCAS 2000 B LCD display Case cover Snap latch Display Operating switch Power lamp green Alarm2 lamp red Alarm1 lamp red sample water out...

Page 6: ...show conditions applying current All terminal blocks are removable 2 3 2 Terminal block and FOCAS 2000 inner connection Operation record memory Backup battery for the clock Reset switch Desiccant sil...

Page 7: ...ay terminals Applied voltage output Specification A Specification B 1 3 Alarm2 Alarm signal 2 0 No output Standby Standby During operation With supply 0 OFF ON OFF OFF ON OFF ON Blinking 4 2 2 OFF OFF...

Page 8: ...water which is indispensable for sensor correction In order to prevent dust contamination from fresh water line please be sure to carry out cleaning of the line Tubes to be connected to the sample wat...

Page 9: ...es stable for measurement The bilge cannot be drained independently of where the oil alarm point exists Example of display during operation Oil concentration Alarm1 oil alarm point Current hours minut...

Page 10: ...before using the oil concentration detector If not the oil concentration detector cannot run When the clock is not set an alarm is output The alarm is reset when the clock is set If the oil concentra...

Page 11: ...ks at a digit you can enter a value Press to select a value Press Enter after selecting a value Repeat the above to set hours minutes and seconds NQEM WUVOGPV AAAA 06A ZZ ZZ ZZAAZZ ZZ ZZ Set seconds a...

Page 12: ...on is less than the specified oil alarm point during manual operation Manual test operation procedure 5VCPFD ZZ ZZ ZZ ZZ ZZ ZZ Make sure the oil concentration detector is in the standby state Standby...

Page 13: ...he alarm test About 5 minuites later with Esc unpressed sim disappears from the screen The operation ends Main Menu returns And then the oil concentration detector enters the standby state sim disappe...

Page 14: ...eration This screen shows the state of the three way valve output signal and the measured oil concentration 1RGTCVKPI UKO AZZ ZZ ZZ 8CNXG 10 A ZZARRO When the oil concentration is less than the oil al...

Page 15: ...easured Clean the sensor with brush see 8 4 to maintain the performance of the detector It is necessary to temporarily cancel the Warning state to enable sensor cleaning with brush see 8 4 Press Enter...

Page 16: ...2 seconds or more Wrong operation prevention Return to standby During standby a clock time is displayed as shown below All data Status of the power supply source Operational status Oil alarming statu...

Page 17: ...ss Enter to set a year The cursor blinks at a digit you can enter a value Press to select a value Press Enter after selecting a value NQEM WUVOGPV 06A ZZ ZZ ZZAAZZ ZZ ZZ Press The cursor moves to a mo...

Page 18: ...ent 20 to 24 1 20 mA During setting Maint Current 0 or 20 to 24 1 0 mA RECout With equipment fault Defect Current 0 or 20 to 24 1 0 mA Setting method Standby 5VCPFD ZZ ZZ ZZAAZZ ZZ ZZ Keep on pressing...

Page 19: ......

Page 20: ...ration Zero point check from the upper lid Zero pointcheck from the fresh water line and RECout output test See 6 1 Tabulated list of menu items To execute a testing function keep on pressing Esc for...

Page 21: ...goes to washing process The remaining time in seconds is displayed to the right of Washing When the washing process ends the oil concentration detector waits for judgment of cleanliness of the detect...

Page 22: ...the sensor unit and open the drain valve once more to empty the sensor unit Press Enter KNN OKPGTCN YCVGT KPVQ VJG EGNN CPF RTGUU GPVGT Repeat a set of feeding mineral water into the sensor unit and d...

Page 23: ...water through a fresh water line on a ship 5VCPFD ZZ ZZ ZZAAZZ ZZ ZZ Keep on pressing Esc for two seconds or longer CKP GPW NQEM FLWUVOGPV Press twice CKP GPW 6 56 QFG Press Enter and then press twic...

Page 24: ...1 OnlyAlarm1 is ON alarm state Relay contact check 2 OnlyAlarm2 is ON alarm state Relay contact check 3 BothAlarm1 andAlarm2 are ON alarm state In the alarm state AL1 andAL2 alarm lamps blink accordin...

Page 25: ...ON Open Page 3 1 2 3 and 4 Level values for each sensor Temp Sensor temperature each sensor level and temperature value are checked while FOCAS 2000 is measuring oil content in sample water or fresh w...

Page 26: ...d drain the spent water from the detection tube and use this function again to feed water for a preset time period to wash the sensor and the detection tube Manual washing procedure Check and make sur...

Page 27: ...ompensation The upper lid and drain valves are opened and closed for compensation Firmly close them when closing them 5VCPFD ZZ ZZ ZZ ZZ ZZ ZZ Keep on pressing Esc for two seconds or longer CKP GPW NQ...

Page 28: ...ration over 5 ppm is detected in the zero point compensation GTQ EQOR 9CTPKPI EGNN VQQ KTV Displayed when the inside of the detection tube is dirty In this display example 10 ppm o 10 ppm To warn that...

Page 29: ...d group Groups related toAutomatic operation Manual opration Washing etc Press 1RGTCVKQPCN QI 1KN NCTO ZZZ Alarm1 andAlarm2 related group Group related to oil alarm points preset oil concentration val...

Page 30: ...er Page 2 of details is displayed Detailed display Page 2 A ZZ ZZ ZZAZZ ZZ ZZ 8 ZZA964 ZZA5 2 ZZA2 ZZ Fresh water valve state xx OFF Close ON Open Valve state xx OFF Close ON Open Pressure switch stat...

Page 31: ...right part of the screen You can also view old records by pressing on this screen Press Enter Page 2 ofAlarm Setting Details is displayed Detailed display of Alarm Setting Page 2 A ZZ ZZ ZZAZZ ZZ ZZ...

Page 32: ...rainport drainwaterfromthedetectiontube andclosethedrainport Feedmineralwaterto80 ofthedetectiontube openthedrainport drainwaterfromthesamplewater detectiontube checkandmakesurethedrainsealwasherisfit...

Page 33: ...ive battery is left exhausted in the oil concentration detector the clock may be reset when power to the oil concentration detector is shut down and Warning 17C is displayed on the screen At the same...

Page 34: ...accordance with 2 2 2 1 of 2013VGP Management and storage of the operational log memory card In some cases the operational log memory card is required for inspection according to rules Put the replace...

Page 35: ...connection When the cable connection is right check cables for disconnection or damage Replace the cable if defective Inspect the connected measuring instruments Check whether the terminal operation...

Page 36: ...n the oil concentration detector can run normally When the error still occurs turn off the oil concentration detector and replace the sensor unit Failure of the light source system in the sensor of th...

Page 37: ...ump pressure source and the power supply source TTQT JGEM UGPUQT RQYGT ECDNG The connectors of the sensor unit are disconnected Confirm the connection of Connector 2 according to 2 3 Internal layout f...

Page 38: ...concentration detector is a high precision instrument capable of detecting oil in parts per million To prevent the product against failures and maintain its accuracy the product must be calibrated wit...

Page 39: ...FF AL 1 62 Alarm 2 ON 63 Alarm 2 OFF AL 2 64 Oil over 30ppm 65 Oil under 30ppm 30ppm Defect 0A Maintenance mode 2A ret fm Maintenance mode 01 Over Range 21 ret fm Over Range 02 Sensor Down 22 ret fm S...

Page 40: ...djust done Zero 88 Performance check 89 Performance check done 8A Accuracy check 8B Accuracy check done 8C Exam mineral water zero 8D Exam mineral water done Zero 8E Exam clean water zero 8F Exam clea...

Reviews: