To change, use the Up/Down arrows to cycle through the mode
options. Press the Left arrow to exit to the main menu.
Setup > Set Zero
The set zero process sets a new baseline, or zero, for the
sensors:
C2H4
CO2
O2
Set zero should take place weekly for the Ethylene (C2H4),
Oxygen (O2) sensors and Carbon Dioxide (CO2) sensors to
maintain maximum performance. To set zero, first press the
right arrow to select the sensor. The set zero process requires
100% nitrogen gas to set zero for O2, an internal Potassium
Permanganate jar to set zero for C2H4, and an option to use an
external conditioning chamber full of Soda Lime or nitrogen gas
to set zero for CO2. Wait for a beep to verify the set zero
process was completed. Setting zero is important in establishing
a zero baseline for the sensors.
Mode
Measure
Trigger
Connection
USB Storage