System Options
:
Plasma Cleaner
C O N F I D E N T I A L – FEI Limited Rights Data
7-44
Sample Cleaning Procedure
1.
Move the stage to the lowest Z-axis position and tilt to zero.
This is to avoid any possible shadowing effect by the column if the sample is at a short working distance.
2.
Run the
Tools
menu /
Sample Cleaning
procedure.
It is possible to enter the procedure from the vented state with the use of
Sample Exchange
window also. During
the procedure progress, stage moves are disabled temporarily.
When the procedure is finished or aborted by a user, the system remains pumped.
3.
Move stage back to the observation position.
When starting the procedure and some conditions are not allowed, the
dialog or tooltip appears onscreen.
It is possible to start cleaning procedure by ticking the Vacuum
module / Sample Cleaning check box (duration can be set) at the
same tim with pumping the chamber to the desired vacuum
mode.
Note
When one is going to observe in LoVac immediately after plasma
cleaning with a different gaseous environment, it is advisable to run
the Purging procedure (see Chapter 3).
Chamber Cleaning
To periodically clean the whole system (due to build ups from samples, carriers, service events, etc.) use the
Chamber Cleaning procedure from the
Plasma Cleaning
Alignment. The clean chamber environment enables an
operator to use the system at a level of cleanliness suitable for high resolution imaging and analysis.
C a u t i o n !
Never use Plasma Cleaner for Chamber Cleaning with CryoCleaner filled with LN2!
W A R N I N G !
If EDS / WDS / EBSD system is mounted, use 1 mm collimator on EDS detector and do not exceed 5 minutes overall
procedure duration once a day at most.
Always retract EDS detector before plasma cleaning!
Do not exceed 30 minutes of chamber cleaning once a week at most when the DBS detector is installed.
Avoid leaving sensitive carbon containing samples (e.g. photo resist) inside the chamber during Plasma Cleaning
procedure, as they may be etched by the cleaning process.
Any material that can create or release oxide easily (e.g. silver) should not be plasma cleaned.
The Au-C resolution tests samples can be left inside the chamber during the Sample Cleaning procedure only and not
very often!
Summary of Contents for Scios 2
Page 1: ...User Operation Manual Edition 1 Mar 2017 ...
Page 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Page 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Page 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Page 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...