Microscope PC’s Desktop Sharing
Several users can connect to the Microscope PC at the same time and thus share its desktop on the remote PC's.
This can be used e.g. for the educational purposes. The typical scenario is that a user controls the microscope
(locally or remotely) and other connected users only observes the Microscope PC's screen over the network.
1.
Start the VNC application
2.
Open the VNC Viewer application on the remote PC of the person who is intended to control the microscope
and connect remotely to the Microscope PC as user.
Note
The desktop sharing must be enabled in the VNC Server application on the Microscope PC. For instructions how to do so,
see VNC Server documentation available on the installation CD.
Controlling Microscope Remotely
After connecting remotely to the Microscope PC, you can do all the tasks as if you work directly with that
computer. It is possible to control the microscope operation, run the supportive applications, backup or restore
user data, view the log files etc.
In general, expect slower response on your commands and actions. This limitation comes from the amount of data
that have to be transmitted and from the network bandwidth.
In case the Microscope Control user interface stops responding or the
Server Busy
dialog appears, the Microscope
PC is probably overloaded and may have problems with the actual conditions being operated remotely. If such a
problem occurs, wait till the Microscope PC recovers; or restart it yourself when necessary.
Here are some recommendations for the remote control operation:
•
To obtain better remote performance, try to use the
View
menu /
Remote Mode
item (see Chapter 3).
•
Be very careful when navigating the stage. Always keep in mind that what you see in the optical display is not
the exact position of the stage (because of delay in imaging data transmission).
•
You can improve the performance / response time by setting the more convenient scanning conditions: the
quad image instead of the single image mode, single display running instead of more displays active; lower
resolution; slower scan (i.e. higher dwell time).
•
Benefit from the automatic functions that are not affected by the delayed response problem (Auto Contrast
Brightness, Auto Focus, Auto Stigmator). In case you prefer to focus or correct astigmatism manually, it is better
to do so in the reduced area.
UMB Stub Holder Kit
This option contains the following items:
•
Universal mounting base (UMB)
•
Storage box
•
2 stub modules each holding 3× 1/2” stubs or 2× 1” stubs
•
2 clamp bars
•
User guide
•
System calibration sample
The stub holder is mounted on the stage interface plate.
Summary of Contents for Scios 2
Page 1: ...User Operation Manual Edition 1 Mar 2017 ...
Page 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Page 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Page 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Page 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...