System Operation:
Vacuum Statuses
3-3
Vacuum Statuses
The vacuum status controls are in the
Vacuum
module. The
Pump
button starts pumping for the operating pressure and the
Vent
button
starts venting for a sample or detector exchange.
In the
Status
module at the bottom of any page the actual vacuum
status is represented by the colored icon, which may have three
possible colors with the following meaning:
•
Green: PUMPED
to the desired vacuum mode
•
Orange: TRANSITION
between two vacuum modes
(pumping / venting / purging)
•
Grey: VENTED
for sample or detector exchange
PUMP BUTTON
When the
Pump
button is clicked and the status is
Vented,
or when
changing vacuum mode, the target pressure that the system pumps to
depends on the selected vacuum mode. The Pump button is
highlighted in and not accessible.
For the
High Vacuum
the system achieves the lowest pressure
possible. For the
Low Vacuum
or
ESEM
it achieves the pressure
specified in the
Vacuum
module
Chamber Pressure
adjuster. The
purge function can be defined in the
Preferences… / ESEM
dialogue.
When the
Pump
button is clicked and the status is
Transition
(venting), the venting procedure stops and the system immediately
starts to pump to the actually selected vacuum mode.
VENT BUTTON
When the
Vent
button is clicked and the status is
Vacuum
, the
confirmation dialogue appears. After confirmation, the system
switches off the detectors voltages, high voltage supplies, vacuum
pumps and uses the appropriate valves to vent the system, with the
use of an air or a gas (the dry Nitrogen typically) brought to the
Nitrogen Inlet
. The Nitrogen is recommended to obtain lower
pressure for high resolution imaging in HiVac mode.
The Vent button is highlighted in and not accessible. After a specified
venting time the venting valve closes and the vacuum status
should
indicate
Vented
. The button is enabled again.
When the
Vent
button is clicked and the status is
Transition
(pumping), the dialogue appears. After confirmation, the pumping
procedure stops and the venting procedure starts.
When the
Vent
button is clicked and the status is
Vented
, the
dialogue appears. After confirmation, the venting valves re-open for
the specified venting time and then the valves close.
Summary of Contents for Quanta FEG 250
Page 8: ...C vi...
Page 28: ...System Operation Quanta FEG System States 3 10...
Page 108: ...Alignments 1 Gun Alignment 6 6...
Page 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Page 152: ...System Options Automatic Aperture System 9 4...
Page 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Page 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...