Stages:
Stage Related Functions
7-14
SPECIMEN ALIGNMENT
xT Align Feature
This utility is designed specifically for long features, extending off the
screen at the magnification required for an observation. It applies the
mapping process bringing the long feature either to the horizontal or
vertical axis to make the navigation easier. This can be performed at
any point within the stage field limits and takes into account the stage
rotation offset.
FIGURE 7-12
XT ALIGN FEATURE
Note:
xT Align Feature works best at the eucentric position (see above).
Longer distances result in a higher accuracy.
C a u t i o n !
Watch the obstacles significantly extending from the sample plane, as
these may interfere with equipment under the lens.
1.
Select a long feature of interest on the sample.
2.
Click the
Stage
menu /
xT Align Feature…
Choose either
Horizontal
or
Vertical
, which relates to the desired final sample
orientation. Click the first point along the feature, the
P1
coordinates update.
3.
Click the second point along the feature, the
P2
coordinates
update. Right-clicking anywhere in the imaging area deletes
points, enabling to define them again.
Summary of Contents for Quanta FEG 250
Page 8: ...C vi...
Page 28: ...System Operation Quanta FEG System States 3 10...
Page 108: ...Alignments 1 Gun Alignment 6 6...
Page 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Page 152: ...System Options Automatic Aperture System 9 4...
Page 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Page 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...