Alignments:
3 - Stage Rotation Centre
6-8
3 - Stage Rotation Centre
The stage rotation has a mechanical center and it can be controlled by
changing the
Stage
module /
Coordinates
tab /
R
value. This moves
the stage around its mechanical center. In some circumstances this is
not desired because a rotation around the field of view center would
be more useful.
The following correction is the alignment procedure for the
Compucentric Rotation function. The X / Y offset is calculated so that
the Compucentric Rotation is correct from that point on.
1.
The magnification should be from 500× to 2000×, and the sample
should have a recognizable feature close to the center of the stub,
which is mounted in the stage center. Make sure the tilt is zero.
2.
The stage automatically rotates the view 180º.
Summary of Contents for Quanta FEG 250
Page 8: ...C vi...
Page 28: ...System Operation Quanta FEG System States 3 10...
Page 108: ...Alignments 1 Gun Alignment 6 6...
Page 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Page 152: ...System Options Automatic Aperture System 9 4...
Page 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Page 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...