![FCI ST100A Series Installation, Operation And Maintenance Manual Download Page 111](http://html.mh-extra.com/html/fci/st100a-series/st100a-series_installation-operation-and-maintenance-manual_547363111.webp)
ST100A Series Flow Meter
TROUBLESHOOTING
Fluid Components International LLC
103
Table 34 – Detailed CORE Fault Register (CORE 2V Command)
Octet-Bit
Status Description
Fault Type
Hex Bit Map
0-0
Device Error: If any of these errors occurs: I2C error, UART error, Mutex error, watchdog reset
Fatal
0x000000000001
0-1
FCI Reserved (Not Used)
Non-Fatal 0x000000000002
0-2
FCI Reserved (Not Used)
Non-Fatal 0x000000000004
0-3
CORE unable to update process data (PD_NO_FE_DATA). Unable to obtain/use data from any Active FEs
Fatal
0x000000000008
0-4
FCI Reserved (Not Used)
Non-Fatal 0x000000000010
0-5
CORE detects FRAM/SPI error
Fatal
0x000000000020
0-6
CORE reports SD card error. Either initialization (corrupt card) error, or card became full (error while writing). For ST100A only.
Non-Fatal 0x000000000040
0-7
FCI Reserved (Not Used)
Non-Fatal 0x000000000080
1-0
FCI Reserved (Not Used)
Non-Fatal 0x000000000100
1-1
FCI Reserved (Not Used)
Non-Fatal 0x000000000200
1-2
FCI Reserved (Not Used)
Non-Fatal 0x000000000400
1-3
FCI Reserved (Not Used)
Non-Fatal 0x000000000800
1-4
FCI Reserved (Not Used)
Non-Fatal 0x000000001000
1-5
CORE unable to communicate with one or more FEs (PD_COMM_ERROR)
Fatal
0x000000002000
1-6
FCI Reserved (Not Used)
Non-Fatal 0x000000004000
1-7
CORE: averaged flow out of range of “Flow Min” or “Flow Max”
Non-Fatal 0x000000008000
2-0
FCI Reserved (Not Used)
Non-Fatal 0x000000010000
2-1
FCI Reserved (Not Used)
Non-Fatal 0x000000020000
2-2
FCI Reserved (Not Used)
Non-Fatal 0x000000040000
2-3
CORE: averaged temperature above “Temperature Max”
Fatal
0x000000080000
2-4
CORE: averaged temperature below “Temperature Min”
Fatal
0x000000100000
2-5
FE Heater is shorted or below operating limit.
Fatal
0x000000200000
2-6
FE Heater monitoring ADC fails to respond.
Fatal
0x000000400000
2-7
FE Heater is open or above operating limit.
Fatal
0x000000800000
3-0
The Reference-R ADC converter fails to respond.
Fatal
0x000001000000
3-1
FCI Reserved (Not Used)
Non-Fatal 0x000002000000
3-2
The Reference-R ADC’s count number is below the minimum.
Non-Fatal 0x000004000000
3-3
FCI Reserved (Not Used)
Non-Fatal 0x000008000000
3-4
Process flow is above the maximum limit.
Non-Fatal 0x000010000000
3-5
FCI Reserved (Not Used)
Non-Fatal 0x000020000000
3-6
FCI Reserved (Not Used)
Non-Fatal 0x000040000000
3-7
The ADC for monitoring the temperature inside the unit fails to respond.
Non-Fatal 0x000080000000
4-0
The Delta-R ADC fails to respond.
Fatal
0x000100000000
4-1
ST80: Not used. ST100A: 4-20mA Input / Pressure Input ADC fails to respond.
Non-Fatal 0x000200000000
4-2
Reference-R value is above operating limit.
Non-Fatal 0x000400000000
4-3
The Inter-Integrated Circuit channel 0 bus fails to communicate.
Non-Fatal 0x000800000000
4-4
The ADC for monitoring heater’s conditions fails to respond.
Fatal
0x00100000000
4-5
The port expansion integrated chip fails to respond.
Non-Fatal 0x00200000000
4-6
FCI Reserved (Not Used)
Non-Fatal 0x00400000000
4-7
FCI Reserved (Not Used)
Non-Fatal 0x00800000000
5-0
The FRAM of the Flow Element fails to respond.
Non-Fatal 0x01000000000
5-1
The Active excitation current Integrated Circuit (IC) fails.
Non-Fatal 0x02000000000
5-2
The Reference excitation current Integrated Circuit (IC) fails.
Non-Fatal 0x04000000000
5-3
CORE: process data are not updated because (all) FE's are in self-test mode, or the system is in non-operating mode.
Non-Fatal 0x08000000000
5-4
Reference-R value is below operating limit.
Non-Fatal 0x10000000000
5-5
Delta-R value is above operating limit.
Non-Fatal 0x20000000000
5-6
Delta-R value is below operating limit.
Non-Fatal 0x40000000000
5-7
The unit is performing a (diagnostic) test.
Non-Fatal 0x80000000000
Notes: 1. Fault Type reflects factory default programming. The Fatal/Non-Fatal designation is programmable via the Configurator software’s FE Faults tab (Factory
branch menu tree – requires entry of appropriate level password).
Summary of Contents for ST100A Series
Page 1: ......
Page 115: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 107 ...
Page 116: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 108 Fluid Components International LLC ...
Page 117: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 109 ...
Page 118: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 110 Fluid Components International LLC ...
Page 119: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 111 ...
Page 120: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 112 Fluid Components International LLC ...
Page 121: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 113 ...
Page 122: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 114 Fluid Components International LLC ...
Page 123: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 115 ...
Page 124: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 116 Fluid Components International LLC ...
Page 125: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 117 ...
Page 126: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 118 Fluid Components International LLC ...
Page 127: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 119 ...
Page 128: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 120 Fluid Components International LLC ...
Page 129: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 121 ...
Page 130: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 122 Fluid Components International LLC ...
Page 131: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 123 ...
Page 132: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 124 Fluid Components International LLC ...
Page 133: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 125 ...
Page 134: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 126 Fluid Components International LLC ...
Page 135: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 127 ...
Page 136: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 128 Fluid Components International LLC ...
Page 137: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 129 ...
Page 138: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 130 Fluid Components International LLC ...
Page 139: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 131 ...
Page 140: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 132 Fluid Components International LLC ...
Page 141: ...ST100A Series Flow Meter APPENDIX A DRAWINGS Fluid Components International LLC 133 ...
Page 142: ...APPENDIX A DRAWINGS ST100A Series Flow Meter 134 Fluid Components International LLC ...
Page 167: ...ST100A Series Flow Meter APPENDIX E CUSTOMER SERVICE Fluid Components International LLC 159 ...
Page 168: ...APPENDIX E CUSTOMER SERVICE ST100A Series Flow Meter 160 Fluid Components International LLC ...