FS10 Series
OPERATION
Fluid Components International LLC
45
Password Protection
Factory settings are protected with a level 1 factory password. End user settings are protected with a level 2 password. The following
password allows changes to any level 2 parameter:
19113
Refer to “Table 9 – Serial Interface CLI Parameter Reference Table” on page 46 for the CLI parameter reference. Access to all level 2 (user
configurable) parameters are available through the FC10 Windows PC interface software.
Command Line Interface (CLI) Commands
The table below summarizes the FS10 command line interface (CLI) command set.To be recognized by the system, each command must start
with an asterisk followed by the command mnemonic plus any applicable argument as specified in the table. The CLI commands can be
entered either as upper or lower case letters (it doesn’t matter which). Commands typically respond with parameter value(s) each identified by
an item number (for example #220–#238 shown in example screen at left for
*meas
command). All possible parameters are listed in “Table
9 – Serial Interface CLI Parameter Reference Table” on page 46.
Table 8 – CLI Command Set
Command Mnemonic
Description
*INFO
Read version information (Hardware, Firmware, Date code, other info).
The parameter values for items #1–#66 are listed.
*MEAS
Read most recent flow measurement values. The parameter values for
items #220 –#238 are listed.
*RCFG B
Read configuration parameters from bank B (0–9). The parameter val-
ues for items #80–#133 are listed for the specified bank number.
Example: *RCFG 7 (to read bank 7 parameters)
*SAVE B
Save currently active parameters to bank B (0–9) (Level 2 – Field pass-
word required).
Example: *SAVE 7
*RCL B
Recall parameters from bank B as the active parameters. (Level 2 –
Field password required).
Example: *RCL 7
*PASSWD NNNN
To enter the password for either Level 1 (Factory) or Level 2 (Field).
Example: *PASSWD 19113
*EXIT
To exit the current password level (undo having entered a password).
The current parameters will automatically be saved to Bank 0 upon exit.
*NNN=<VALUE>
Update item #NNN (#1–#238) in the “active” (RAM) parameters with
value. Value may be either integer or floating point (scientific notation).
Example 1: *228=1.03 enters value 1.03 into item #228
Example 2: *1=002 enters 002 into item #1
Example 3: *110=800 enters 800 into item #110
*B:NNN
For Bank B (0–9) read item #NNN (#1–#238). For items #1–#45, and items
#220–#238 the Bank is ignored. If no bank is specified, the currently ac-
tive (RAM) parameters will be accessed.
Example 1: *7:85 returns item 85 from bank 7 in this form:
7:85>5.053665E-02
Example 2: *8:119 returns item 119 from bank 8 in this form: 8:119=030]