FaroArm
®
Quantum
August 2017
79
Chapter 3: Operation
Diagnostics
The
Diagnostics
group controls is a set of commands for checking your
Quantum’s setup and general operation.
Base Deflection
Click B
ASE
D
EFLECTION
to run a test the physical setup of your Quantum. If this
test fails, run Diagnostic Tilt command to test for any tilt in the base of the
Quantum.
1.
Set up the FARO compensation cone approximately
⅔
of the Quantum's
reach away from its base.
2.
Place a properly compensated probe firmly in the cone.
3.
Click
Reset Data
to begin the test.
4.
Press the green
Front
button and move the tubes and joints of the Quantum
while keeping the probe firmly seated in the cone.
N
OTE
: Press the green
Front
button to reset the screen at any time.
5.
Press the red
Back
button to stop.
Figure 3-47 Network Configuration dialog box
08m80e00_FaroArm_Quantum.book Page 79 Tuesday, August 22, 2017 11:03 AM
Summary of Contents for FaroArm Quantum
Page 2: ...08m80e00_FaroArm_Quantum book Page 2 Tuesday August 22 2017 11 03 AM...
Page 4: ...08m80e00_FaroArm_Quantum book Page 2 Tuesday August 22 2017 11 03 AM...
Page 10: ...08m80e00_FaroArm_Quantum book Page vi Tuesday August 22 2017 11 03 AM...
Page 128: ...08m80e00_FaroArm_Quantum book Page 118 Tuesday August 22 2017 11 03 AM...
Page 154: ...08m80e00_FaroArm_Quantum book Page 144 Tuesday August 22 2017 11 03 AM...
Page 170: ...08m80e00_FaroArm_Quantum book Page 160 Tuesday August 22 2017 11 03 AM...
Page 174: ...164 08m80e00_FaroArm_Quantum book Page 164 Tuesday August 22 2017 11 03 AM...
Page 190: ...08m80e00_FaroArm_Quantum book Page 6 Tuesday August 22 2017 11 03 AM...
Page 200: ...08m80e00_FaroArm_Quantum book Page 6 Tuesday August 22 2017 11 03 AM...
Page 201: ...08m80e00_FaroArm_Quantum book Page 1 Tuesday August 22 2017 11 03 AM...