Cement Curing Autoclave User Manual
100072577
Revision H
36
5.5.2
Coat the chamber threads with grease.
Figure 5-13
Grease chamber threads
5.5.3
Fill the chamber with water to halfway up the threads. This ensures
all air will be removed from the chamber when the lid is installed. See
API RP 10B.
5.5.4
Inspect the chamber lid O-ring for cuts or other damage. If it is
damaged, it should be replaced.
5.5.5
Grease the lid O-ring and lid threads.
5.5.6
Install the lid onto the chamber. Fully tighten the lid into the
chamber using the pressure vessel wrench and Vise Grip chain clamp.
When the lid is fully seated, loosen it 1/8 to 1/4 turn. This helps to ensure
the lid does not become jammed.