C=20log(Vs/Vn)=6.4dB
S: Overall Sensitivity
S=A+B+C=97.1db
Supplementary information: --
Distance for the probe face
100mm
Nominal acoustic working frequency
3.0MHz
Doppler frequency
293Hz
Target Velocity
12cm/s
A(d): Target reflection loss
44.5dB
B: two-way attenuation over the acoustic
pathway
46.8dB
C: Signal-to-noise Ratio
Vs: 227mV
Vn: 109mB
C=20log(Vs/Vn)=6.4dB
S: Overall Sensitivity
S=A+B+C=97.7db
Supplementary information: --
Distance for the probe face
200mm
Nominal acoustic working frequency
3.0MHz
Doppler frequency
293Hz
Target Velocity
12cm/s
A(d): Target reflection loss
44.5dB
B: two-way attenuation over the acoustic
pathway
41.4dB
C: Signal-to-noise Ratio
Vs: 225mV
Vn: 106mB
C=20log(Vs/Vn)=6.5dB
S: Overall Sensitivity
S=A+B+C=92.4db
Supplementary information: --