background image

 

- 0 - 

 
 
 
 
 
 
 
 

 

 

MS 6000 

BED SCALE  

                    USER MANUAL 

Please keep the instruction manual at hand all the time for future reference.

 

 

Summary of Contents for Charder MS 6000

Page 1: ... 0 MS 6000 BED SCALE USER MANUAL Please keep the instruction manual at hand all the time for future reference ...

Page 2: ... 14 Preparation for Measurements 15 Installing Weight Bridge 17 Panel 18 Key Functions 18 LCD Symbol Description 19 Date Time Setting Procedure 20 Normal Weighing Procedure patient on the bed 21 Weight Tracking Procedure 22 Using the Bed Scale 24 Using the BMI Feature 25 Using the Tare Function 26 Storing and Re Calling PRE TARE Weight 26 Set Up your Device 29 Printing Function 30 Instructions for...

Page 3: ...the Operating Instructions They contain important instructions for installation proper use and maintenance of the device The manufacturer shall not be liable for damages arising out of failure to heed the following instructions These batteries should be kept away from small children If swallowed promptly seek medical assistance Expected Service Life 5 years When using electrical components under i...

Page 4: ...cted Cleaning We would recommend using alcohol based wipes or similar when cleaning the scales Please do not use large amounts of water when cleaning the scales as this will cause damage to the scales electronics you should also refrain from using corrosive liquids or high pressure washers Always disconnect the scales from the mains power supply before cleaning Maintenance The scale does not requi...

Page 5: ...e property of Faaborg Rehab Should the fault repairs or replacement delivery not be successful the statutory provisions shall be valid The period of warranty shall be two years beginning on the date of purchase Please retain your receipt as proof of purchase Should your scale require servicing please contact your dealer or Faaborg Rehab s Customer Service No responsibility shall be accepted for da...

Page 6: ...faults for the length of the warranty period Disposing of the Scale This product is not to be treated as regular household waste but should be handed into an electrical electronic equipment recycling centre You can obtain further details from your local council your municipal waste disposal company or the firm which you purchased the product ...

Page 7: ...te the accompanying documents or Observe operating instructions Charder Electronic Co Ltd No 103 Guozhong Rd Dali Dist Taichung City 412 Taiwan R O C Identification of manufacturer of medical product including address Type B applied part Dispose of old appliances separately from your household waste l Instead take them to communal collection points ...

Page 8: ...on manual before setup and commissioning even if you are already familiar with Charder scales Transport and storage temperature limit indicating the upper and the lower limit Transport and storage temperature on packaging 60 C 20 C ...

Page 9: ...RF emissions CISPR 11 Group 1 The MEDICAL SCALE MS6000 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The MEDICAL SCALE MS6000 is suitable for use in all establishments including domestic establishments and those directly connected to the public low volta...

Page 10: ... applicable Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV line s to line s 2kV line s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5...

Page 11: ...S6000 should assure that is used in such and environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 3 Vrms 150 KHz to 80 MHz 3 Vrms Portable and mobile RF communications equipment should be used no closer to any part of the MEDICAL SCALE MS6000 including cables than the recommended separation distance calculated from the equa...

Page 12: ...d transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the MEDICAL SCALE...

Page 13: ...ransmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to t...

Page 14: ...on 109 X 7 5 X 25 5 cm Weighing Area 88 X 18 cm Key Functions ON OFF ZERO HOLD BMI PRINT PRE TARE CLEAR ENTER M1 10 MONITOR Weight ALARM 0 9 KEYS Power Supply 6 AA size batteries Rechargeable battery pack optional 12V 2A switching adaptor Parts overview Weigh bridge x 2 connecting wire x 1 DP3700 Indicator x 1 Adaptor x 1 User manual x 1 RS232 cable Bracket set for wall mount x 1 Operation Tempera...

Page 15: ... 14 Power Adapter Standards CAUTION The device is only compatible with restricted power adaptors in dashed block below ...

Page 16: ...g weigh bridges and power on the scale before pushing the bed on it Please refer to installing instruction 4 Set the unit on a stable level surface When the bubble indicator is in the centre of the circle MS6000 is leveled correctly Bubble indicator Good Wrong All four feet must touch the surface to make sure that MS6000 does not rock ...

Page 17: ...trongly recommend two people to operate the procedure Before moving bed on the weigh bridge please turn on the scale You must make sure that the connecting cable is not caught under the measuring foot otherwise faulty measurements may result 6 Four bed casters are now placed well on weigh bridges ...

Page 18: ...g wire and Indicator Connect both the weigh bridges using Weigh Bridge connecting wire and then connect Indicator with Weigh Bridge 2 Weigh Bridge 1 Weigh Bridge 2 Note Please connect all the Pins carefully with gentle hands and match with and push the wire connecting slot gently Female plug Female plug Male plug female plug Connecting wire ...

Page 19: ...TARE value PRE TARE TARE PRE TARE function for subtraction of a known bed weight or any other object before weighing process To Tare any unwanted weight HOLD BMI To weigh the patients who are unstable and to lock the weighing value on screen Long press for 3 seconds to calculate the BMI of patient PRINT To print out the results ...

Page 20: ... Alarm function CLEAR To clear the wrong entry while entering digits ENTER To confirm apply 0 9 Entering digits LCD Symbol Description Definitions Stable symbol To indicate that the weight is stable Minor weight Weight under zero Zero symbols Weight is at zero point Low battery Battery need to charge or replaced ...

Page 21: ... To input Dec 25 2008 8 00am Enter year using 0 9 keys Press key after successful entry to move to next Enter date Ex 12 25 enter 12 25 using 0 9 keys Press key after successful entry to move to next Enter time using 0 9 keys and then again press key Display Format YYYY MM DD TT SS First Year then Date Month and then Time will display and then the scale will automatically come to weighing mode ...

Page 22: ... and the minus sign will appear in front of Pre Tare value Place both the weigh bridges under the bed and push the bed on the scale The upper display will show the net weight of the patient Now if you want BMI also simply long press HOLD BMI key for 3 seconds the middle display will show the last height of the patient and extreme digit will blink Key in the height measurement of patient using 0 9 ...

Page 23: ... key Push the bed on the scale the display will show total bed weight on upper LCD Press Pre Tare key to tare the weight of the bed the scale will return to 0 0 reading Move the patient on the bed the upper display will show net weight of patient Long press Monitor Weight key the patient weight will lock and the middle display will display 0 0 reading and the arrow will point to weight ...

Page 24: ...g arrow pointing to sign means the Alarm volume is maximum To decrease the volume level press key again the blinking arrow will stop blinking and to turn off the alarm press the key once the pointing arrow will disappear NOTE As the Weight Tracking Function is activated the original Patient Weight patient weight in Step 4 will lock and saved in memory And this function can be activated only after ...

Page 25: ...l Pre tare the bed weight and sign will appear in front of bed weight Move the bed Patient is in bed on the scale platform Press the HOLD key The HOLD will display on the screen with blinking triangle and after few seconds the average of fluctuating weight will display on the screen and will lock on the display To release the locked weight on display simply press the HOLD key again or remove the o...

Page 26: ...ange as the height value will increase or decrease 4 At this stage the weight and height can be changed and the bed scale will automatically calculate the BMI value according to the changed weight and height 5 HOLD feature can also be used if the weight is unstable 6 Press BMI button to return normal weighing mode Body Mass Index Categories Classification of weight for adults over 18 years on the ...

Page 27: ...the patient need to be weighed on the bed the display will show the net weight of the patient To delete the saved tare value remove the tare object from the tread platform and press PRE TARE key Storing and Re Calling PRE TARE Weight Charder Bed Scale can store 10 sets of pre tare values and there are two methods to store pre tare value Using Dead Weight or Enter weight using 0 9 keys A Using Dead...

Page 28: ...RIPTION EXAMPLE Long press PRE TARE key for 3 seconds the extreme left digit will blink Enter pre tare weight using 0 9 keys For example 2kg Press ENTER key to confirm the pre tare weight the display will show the minus sign on left of the pre tare weight value Press M1 10 key the blinking m sign will appear on the display ...

Page 29: ...d C RECALL PRE TARE VALUE DESCRIPTION EXAMPLE Long press M1 10 key for 3 seconds the display will show pre tare value m1 firstly The pre tare value will blink Press numeral key 0 9 to choose pre tare value Press ENTER key to confirm the pre tare weight the instrument will automatically perform tare function Press PRE TARE key to return to Normal Mode ...

Page 30: ...select the auto turn OFF time of the device BUZZER ON OFF SETUP This enables operator to select beep sound ON OFF Auto off time 120 sec 180 sec 240 sec 300 sec off Buzzer On Off Please Press Pre Tare key to confirm the setting If you want to select as auto off time press Tare key when displays on the screen To apply the changes proceed till END displays on the screen 240 S 240 S ...

Page 31: ...ess PRINT key to print out the results The format presented below is the standard format of results print out and cannot be changed Parameters of RS232 interface Set parameters of the scale interface on the connected device It is not possible to change the scale parameters Baud rate 9600 bps Parity check None Data length 8 bits Stop bit 1 bit Handshake RTS CTS Data code ASCII ...

Page 32: ...Select Your COM Port Click Connect to select your COM port Usually there s only one option for select Then click OK 4 Port Settings Click Bits per second to set up rate at 9600 Data bits at 8 Parity at None Stop bits at 1 and Flow control at Hardware Then click OK to complete your setting 5 Output Data When the patient has been weighed and BMI calculated simply press the PRINT key to output data f...

Page 33: ... the adaptor from main power supply before taking out the AC adaptor pin from indicator Instructions for Printing and Connecting If prompt displays on the LCD please change the scale batteries or use exclusive adaptor to operate the scale Locate adaptor plug on the right side of indicator RS232 Connector AC Jack Connector ...

Page 34: ... 33 Instructions for Replacing Batteries 1 2 3 Open the battery housing cover Take out the battery housing Rechargeable Battery and AA size Battery Housing ...

Page 35: ... 5 AA size Battery Housing Rechargeable Battery Housing Make sure that all the batteries are properly placed Place the battery housing back and make sure that the housing pin is connecting to the right point ...

Page 36: ... 35 6 7 Place the housing back Close the battery housing cover ...

Page 37: ... 36 ERROR Message ...

Page 38: ...p charge if not covered under the warranty Before you contact our Authorized Dealer in your country for technician repair service please read through the following section carefully Self checking Tips Some functional defects can be identified and maintained by users as listed below 1 Power on failure Check if the main power adaptor has not plugged onto the scale properly Check if the battery power...

Page 39: ...hort circuit wires Wire broken Safety fuse burnt out Wire connection problem Main power adaptor faulty Parts Replacement 2 LCD display faulty Possible hardware defects include Uneven brightness in the LCD display screen texts color blurred smeared rainbow screen incorrect decimal display LCD PIN broken or short circuit PCB cooper foil broken loosed welding Unable to save or read data IC or transis...

Page 40: ...ce Directive 2014 31 EU Non automatic Weighing Instruments Directive Please see separate document showing on sticker of device for above CE marking Authorized EU Representative Manufactured by Charder Electronic Co Ltd No 103 Guozhong Rd Dali Dist Taichung City 412 Taiwan R O C FDA no D072479 IN 1224 12185H 11 2016 Faaborg Rehab Technic ApS Smedemestervej 9 DK 5600 Faaborg Denmark Phono no 45 70 2...

Reviews: