8. Multiply the number of divisions counted in Step 7 by the setting of the TIME/DIV
switch to find the measured risetime. If 10X magnification was used, divide the
TIME/DIV setting by 10. For example, if the A timebase setting in Figure 2-11 was 1
s/div(1000ns), the risetime would be 360 nanoseconds (1000ns
÷
10=100ns, 100ns X
3.6 div = 360ns)
9. To Measure falltime, simply shift the trace horizontally until a trailing edge passes
through the 10% and central vertical graticule lines, and repeat Steps 7 and 8.
10. When measuring the rise and fall time, note that 5.8ns-Rise time (tr)=0.35/f-
3dB
which is transition time is contained in the OS-5060A oneself. Therefore the real
transition time (tc) is composed of measure transistion time™ and tr.
The above all is explained with the following formula :
tc=
√
tm
²
- tr
²
tc = Real transition time
tm = Measured transition time
tr = Rise time of oscilloscope
Free Datasheet http://www.datasheet4u.com/
Summary of Contents for OS-5060A
Page 1: ...Oscilloscope Analog Oscilloscope Operation Manual Free Datasheet http www datasheet4u com ...
Page 2: ...Oscilloscope Analog Oscilloscope Operation Manual Free Datasheet http www datasheet4u com ...
Page 13: ...12 Free Datasheet http www datasheet4u com ...
Page 27: ...FIGURE 2 3 USING THE TV SYNC SEPARATOR Free Datasheet http www datasheet4u com ...
Page 30: ...FIGURE 2 4 TRIGGER SLOPE SELECTION Free Datasheet http www datasheet4u com ...
Page 35: ...FIGURE 2 5 SWEEP MAGNIFICATION BY THE B TIME BASE Free Datasheet http www datasheet4u com ...
Page 43: ...FIGURE 2 8 TIME INTERVAL MEASUREMENT Free Datasheet http www datasheet4u com ...
Page 49: ...FIGURE 2 11 RISETIME MEASUREMENT Free Datasheet http www datasheet4u com ...
Page 51: ...4 DIAGRAMS 4 1 EXTERNAL VIEWS Free Datasheet http www datasheet4u com ...
Page 52: ...4 2 BLOCK DIAGRAMS Free Datasheet http www datasheet4u com ...