7
Temp. Coefficient:
Nominal 0.15 x specified accuracy per ºC (between
0 and 18
o
C or 28 to 50
o
C), or as otherwise specified
Power Supply:
9V
battery
(NEDA1604, JIS006P or IEC6F22)
AC Sensing:
True
RMS
Auto Power Off:
After 17 minutes of inactivity with no input signal
Safety:
Intended for indoor use and protected by double insulation per
EN61010-1 and IEC61010-1 2nd Edition (2001) to CAT III
1000V & CAT IV 600V. The meter also meets UL3111-1(1994)*
and CSA C22.2 No. 1010-1-92* to CAT III 1000V.
Terminals (to COM) ratings:
V :
CATEGORY III 1000 VOLTS AC & DC, AND
CATEGORY IV* 600 VOLTS AC & DC.
A / MA
µ
A : CATEGORY III AND CATEGORY IV* 500
VOLTS AC AND 300 VOLTS DC.
*Category IV safety standard (for DMMs) was first released in
IEC61010-1 2nd Edition in year 2001, and was yet a UL published
standard at the time this manual was written.
UL Certification:
Meters that bear the UL marking have been
investigated by UL headquarter in the USA per UL
standard UL3111-1 1st Ed to its highest CAT III rating
and international standard IEC61010-1 Second Edition
(year 2001) to CAT IV rating. The UL markings on the
meter, where applicable, are marked as “UL Listed Cat
III only” and “UL Classified to IEC61010-1 2nd Ed. Cat
IV”.
E.M.C.:
Meets EN61326(1997, 1998/A1), EN61000-4-2(1995) and
EN61000-4-3(1996). Also meets former standards EN55011
(1991) and EN50082-1(1997)
In an RF field of 3V/m: Capacitance function is not specified
Other function ranges: Total Accuracy = Specified Ac 30
digits. Performance > 3V/m is not specified
Overload Protection:
µA/mA Range:
0.63/500V, IR 200kA, F fuse
A Range:
12.5A/500V IR 20kA, F fuse
V Range:
1050V rms, 1450V peak
mV,
Ω
and other:
600VDC/VAC rms
Power Consumption
: 2.7mA
typical
Dimension:
7.32 x 3.43 x 1.4" with holster (186mm x 87mm x 35.5mm)
Weight:
15.17 oz. with holster (430g)