
5
AC CURRENT
RANGE Accuracy
Burden
Voltage
50 / 60Hz
500.0
µ
A, 5000
µ
A
0.6%+3d
0.15mV/
µ
A
50.00mA, 500.0mA
0.6%+3d 3.3mV/mA
5.000A, 10.00A*
0.6%+3d 0.03V/A
40Hz to 1kHz
500.0
µ
A, 5000
µ
A
0.8%+4d
0.15mV/
µ
A
50.00mA 0.8%+4d
3.3mV/mA
500.0mA 1.0%+4d
3.3mV/mA
5.000A, 10.00A*
0.8%+4d 0.03V/A
*10A continuous, 20A for 30-second max with 5 minute cool down interval
TEMPERATURE
RANGE Accuracy
-50
°
C TO 1000
°
C
0.3% + 3d
-58
°
F TO 1832
°
F
0.3% + 6d
RESISTANCE
RANGE Accuracy
50.00
Ω
0.2% + 6d
500.0
Ω
0.1% + 3d
5.000k
Ω
, 50.00k
Ω
, 500.0k
Ω
0.1% + 2d
5.000M
Ω
0.4% + 3d
50.00M
Ω
1.5% + 5d
Open Circuit Voltage: < 1.3VDC (< 3VDC for 50
Ω
& 500
Ω
ranges)
CAPACITANCE
RANGE Accuracy*
50.00nF, 500.0nF
0.8% + 3d
5.000
µ
F
1.0% + 3d
50.00
µ
F
2.0% + 3d
500.0
µ
F
3.5% + 5d
9999
µ
F
5.0% + 5d
*Accuracies with film capacitor or better