Specifications
Specifications
stated
for
ambient
conditions
23
o
C
±5
o
C
(73.4
o
F
±9
o
F);
Relative
Humidity
<
75%
Electrical
Specifications
Function
Range
Resolution
Accuracy
DC
Voltage
40.00mV
10
V
(0.7%
of
reading
+
5
digits)
400.0mV
0.1mV
(0.5%
of
reading
+
2
digits)
4.000V
1mV
(0.7%
of
reading
+
3
digits)
40.00V
10mV
400.0V
0.1V
600V
1V
DCV
Notes:
10M
Input
Impedance
Max
input
voltage
±600V
AC
Voltage
40.00mV
10
V
(1.0%
reading
+
3
digits)
400.0mV
0.1mV
4.000V
1mV
(0.8%
reading
+
3
digits)
40.00V
10mV
400.0V
0.1V
600V
1V
(1.0%
reading
+
3
digits)
VFC
200.0~600V
0.1/1V
(4.0%
reading
+
3
digits)
ACV
Notes:
10M
Input
Impedance
Max
input
voltage
±600Vrms
Frequency
response:
45~1kHz
(for
VFC:
45~400Hz)
True
RMS
All
AC
voltage
ranges
are
specified
from
5%
of
range
to
100%
of
range
Crest
Factor:
≤
3
at
full
scale
up
to
400V,
decreasing
linearly
to
≤
1.5
at
600V
DC
Current
400.0
A
0.1
A
(0.7%
reading
+
2
digits)
4000
A
1
A
40.00mA
10
A
400.0mA
0.1mA
4.000A
1mA
(1.0%
reading
+
3
digits)
10.00A
10mA
DCA
notes:
Overload
protection
for
A/mA
range:
F1
fuse
(
ϕ
6x32)mm;
FF0.5A
H
600V
(CE)
Overload
protection
for
10A
range:
F2
fuse
(
ϕ
6x25)mm
F
10A
H
600V
(CE)
AC
Current
400.0
A
0.1
A
(1.0%
reading
+
3
digits)
4000
A
1
A
40.00mA
10
A
400.0mA
0.1mA
4.000A
1mA
(1.2%
reading
+
3
digits)
10.00A
10mA
ACA
notes:
True
RMS
Frequency
response:
45~1kHz
All
AC
voltage
ranges
are
specified
from
5%
of
range
to
100%
of
range
Overload
protection
same
as
DC
Current
overload
protection
(see
above)
www.
.com