AC CURRENT (FREQUENCY, DUTY CYCLE) MEASUREMENTS 

CAUTION: 

 Do not make 20A current measurements for longer than 30 seconds. 

Exceeding 30 seconds may cause damage to the meter and/or the test leads. 

1. 

Insert the black test lead banana plug into the negative 

COM

 jack. 

2. 

For current measurements up to 4000µA AC, set the function switch to 
the 

µA

 position and insert the red test lead banana plug into the 

µA/mA

 

jack. 

3. 

For current measurements up to 400mA AC, set the function switch to 
the 

mA

 position and insert the red test lead banana plug into the 

µA/mA

 

jack. 

4. 

For current measurements up to 20A AC, set the function switch to the 

10A/HZ/%

 position and insert the red test lead banana plug into the 

10A

 

jack. 

5. Press 

the 

MODE

 button to indicate “

AC

” on the display. 

6. 

Remove power from the circuit under test, then open up the circuit at the 
point where you wish to measure current. 

7. 

Touch the black test probe tip to the neutral side of the circuit.  
Touch the red test probe tip to the “hot” side of the circuit. 

8. 

Apply power to the circuit. 

9. 

Read the current in the display.  In the 10AAC range, right auxiliary display frequency. 

10.  Press and hold the 

MODE

 button to indicate “

Hz

”. 

11.  Read the frequency in the display. 
12.  Momentarily press the 

MODE

 button again to indicate “

%

”. 

13.  Read the % duty cycle in the display. 
14.  Press and hold the 

MODE

 button to return to current measurement. 

10.  With ACA in the main display, press EXIT for 2 seconds to measure AC+DC. 

 

 

 

RESISTANCE MEASUREMENTS 

WARNING:

 To avoid electric shock, disconnect power to the unit under test and 

discharge all capacitors before taking any resistance measurements. Remove the 
batteries and unplug the line cords. 

1. 

Set the function switch to the 

 

CAP

 

 position. 

2. 

Insert the black test lead banana plug into the negative 

COM

 jack. 

Insert the red test lead banana plug into the positive 

Ω

 jack. 

3. Press 

the 

MODE

 button to indicate “

Ω

 

on the display. 

4. 

Touch the test probe tips across the circuit or part under test. It is best 
to disconnect one side of the part under test so the rest of the circuit will 
not interfere with the resistance reading. 

5. 

Read the resistance in the display. 

 
 

 
 
 

Summary of Contents for EX540

Page 1: ...User s Guide Wireless TRMS Multimeter Model EX540...

Page 2: ...more terminals identifies them as being associated with ranges that may in normal use be subjected to particularly hazardous voltages For maximum safety the meter and its test leads should not be han...

Page 3: ...IEC 61010 2 USE EXTREME CAUTION when working with high voltages 3 DO NOT measure voltage if the voltage on the COM input jack exceeds 1000V above earth ground 4 NEVER connect the meter leads across a...

Page 4: ...ty Diode test Battery status n nano 10 9 capacitance micro 10 6 amps cap m milli 10 3 volts amps A Amps k kilo 10 3 ohms F Farads capacitance M mega 10 6 ohms RF transmitter active Ohms PEAK Peak Hold...

Page 5: ...Electrocution The probe tips may not be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets As a result the reading may...

Page 6: ...surements for longer than 30 seconds Exceeding 30 seconds may cause damage to the meter and or the test leads 1 Insert the black test lead banana plug into the negative COM jack 2 For current measurem...

Page 7: ...e red test probe tip to the hot side of the circuit 8 Apply power to the circuit 9 Read the current in the display In the 10AAC range right auxiliary display frequency 10 Press and hold the MODE butto...

Page 8: ...jack and the red test lead banana plug into the positive V jack 3 Press the MODE button to indicate and V on the display 4 Touch the test probes to the diode under test Forward voltage will typically...

Page 9: ...up and connect as described for DC mA measurements 2 Set the rotary function switch to the 4 20mA position 3 The meter will display loop current as a with 0mA 25 4mA 0 20mA 100 and 24mA 125 AUTORANGI...

Page 10: ...1 millisecond in duration Momentarily press the PEAK button PEAK and MAX will display in left auxiliary display MIN will display in right auxiliary display The meter will update the dispay each time...

Page 11: ...he receiver will also blink 5 Once per second the data will be displayed on the pc screen plotted and inserted into a list 6 Hold the backlight button for two seconds to exit the RF wireless transmit...

Page 12: ...Y If it gets wet wipe it off 2 USE AND STORE THE METER IN NORMAL TEMPERATURES Temperature extremes can shorten the life of the electronic parts and distort or melt plastic parts 3 HANDLE THE METER GEN...

Page 13: ...the end user are legally bound EU Battery ordinance to return all used batteries disposal in the household garbage is prohibited You can hand over your used batteries accumulators at collection point...

Page 14: ...the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy...

Page 15: ...0 1V All AC voltage ranges are specified from 5 of range to 100 of range DC Current 400 A 0 01 A 1 0 reading 3 digits 4000 A 0 1 A 40mA 0 001mA 400mA 0 01mA 10A 0 001A 20A 30 sec max with reduced accu...

Page 16: ...0 0001MHz 40MHz 0 001MHz 100MHz 0 01MHz Not specified Sensitivity 0 8V rms min 20 to 80 duty cycle and 100kHz 5Vrms min 20 to 80 duty cycle and 100kHz Frequency electrical 40 00HZ 4KHz 0 01HZ to 0 001...

Page 17: ...ation for positive Minus sign for negative Measurement Rate 2 times per second nominal Low Battery Indication is displayed if battery voltage drops below operating voltage Battery One 9 volt NEDA 1604...

Page 18: ...n any form Garant a Espa ol EXTECH INSTRUMENTS CORPORATION UNA EMPRESA FLIR garantiza este instrumento para estar libre de defectos en partes o mano de obra durante un a o a partir de la fecha de emba...

Page 19: ...nir nos coordonn es Un num ro d autorisation de retour AR doit tre d livr avant tout retour de produit Extech L exp diteur prend sa charge les frais d exp dition le fret l assurance et l emballage cor...

Reviews: