Extech Instruments EX410A User Manual Download Page 13

EX410A-en-GB_V6.0   6/16 

13

REPLACING THE FUSES 

WARNING: 

To avoid electrical shock, disconnect the meter from any circuit, remove the test leads 

from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter 
with an open case. 

1.  Disconnect the test leads from the meter. 

2.  Remove the protective rubber holster. 

3.  Remove the battery cover (two  B  screws) and the 

battery. 

4.  Remove the four  A  screws securing the rear cover. 

5.  Lift the center circuit board straight up from the 

connectors to gain access to the fuse holders.

6.  Gently remove the old fuse and install the new fuse 

into the holder. 

7.  Always use a fuse of the proper size and value (0.2A/600V fast blow (5x20mm) for the 200mA 

range, 10A/600V fast blow (6.3x32mm) for the 10A range). 

8.  Align the center board with the connectors and gently press into place. 

9.  Replace and secure the rear cover, battery and battery cover. 

WARNING

: To avoid electric shock, do not operate your meter until the fuse cover is in place and 

fastened securely. 

www.

.com

[email protected]

1.800.561.8187

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Summary of Contents for EX410A

Page 1: ...USER MANUAL Digital Multimeter MODEL EX410A www com information itm com 1 800 561 8187 www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30...

Page 2: ...or terminal indicates the user must refer to the manual for further information This symbol adjacent to a terminal indicates that under normal use hazardous voltages may be present Double insulation T...

Page 3: ...ical outlets can be difficult and misleading because of the uncertainty of connection to the recessed electrical contacts Other means should be used to ensure that the terminals are not live If the eq...

Page 4: ...0Vrms on 200mV range mA DC 200mA 600V fast acting fuse A DC 10A 600V fast acting fuse 30 seconds max every 15 minutes Ohms Continuity 250Vrms for 15sec max 2 USE EXTREME CAUTION when working with high...

Page 5: ...jack 8 Positive input jack 9 Battery check button 10 Hold button freezes displayed reading 11 LCD backlight button Note Tilt stand test lead holders and battery compartment are on rear of unit Symbols...

Page 6: ...n connected to a circuit DC VOLTAGE MEASUREMENTS CAUTION Do not measure DC voltages if a motor on the circuit is being switched ON or OFF Large voltage surges may occur that can damage the meter 1 Set...

Page 7: ...r V AC positions to obtain a higher resolution reading DC CURRENT MEASUREMENTS CAUTION Do not make current measurements on the 10A scale for longer than 30 seconds Exceeding 30 seconds may cause damag...

Page 8: ...red test probe tip to the hot side of the circuit 6 Apply power to the circuit 7 Read the current in the display RESISTANCE MEASUREMENTS WARNING To avoid electric shock disconnect power to the unit u...

Page 9: ...to the diode under test Forward bias will typically indicate 400 to 1000 Reverse bias will indicate 1 Shorted devices will indicate near 0 and the continuity beeper will sound An open device will indi...

Page 10: ...r the function selected Specifications Function Range Resolution Accuracy DC Voltage V DC 200mV 0 1mV 0 3 reading 2 digits 2V 0 001V 0 5 reading 2 digits 200V 0 1V 600V 1V 0 8 reading 2 digits AC Volt...

Page 11: ...Battery Indication is displayed if battery voltage drops below operating voltage Battery One 9 volt NEDA 1604 battery Fuses mA A ranges 0 2A 600V fast blow A range 10A 600V ceramic fast blow Operating...

Page 12: ...e removed to prevent damage to the unit Battery Replacement 1 Remove the Phillips head screw that secures the rear battery door 2 Open the battery compartment 3 Replace the 9V battery 4 Secure the bat...

Page 13: ...nter circuit board straight up from the connectors to gain access to the fuse holders 6 Gently remove the old fuse and install the new fuse into the holder 7 Always use a fuse of the proper size and v...

Reviews: