2 382860 Ver. 1.5 2/01
2. SPECIFICATIONS
2.1 General Specifications
Display
One Main and three Sub 3-3/4-digit (3999 count) LCDs plus 40-
segment bargraph
Max. Input
Current
DC/AC : 20A
Operating
Conditions
0 to 104
o
F (0 to 40
°
C), relative humidity: <75%, not wetting
Storage
Conditions
0 to 122
o
F (-10 to 50
°
C), relative humidity: <76%, not wetting
Power
9V Battery: NEDA 1604 9 V or 6F22
Dimensions
7.36x3.3x1.34” (187x84x34mm)
Weight
9.8oz (305g)
2.2 Range Specifications
NOTES:
Accuracies are % of reading and specified at 23
o
C with RH <75%
for a period of 1 year after production. Warm-up time is 1 minute.
Mode
Measuring Range
Accuracy
Resolution
400 mV
100
µ
V
4 V
1 mV
40 V
10 mV
400 V
± (0.3% + 1dgt)
100 mV
DC Voltage
1000 V
± (0.5% + 1dgt)
1 V
400
µ
A
0.1
µ
A
4mA
± (1.0% + 1dgt)
1
µ
A
40mA
10
µ
A
400mA
± (0.8% + 1dgt)
100
µ
A
4A
1mA
DC Current
20A
± (1.5% + 5dgt)
10mA
Accuracy
Mode
Measuring Range
40Hz-1KHz
1KHz–10KHz
Resolution
400 mV
0.1mV
4 V
1mV
40 V
10mV
400 V(<200V)
± (0.8%+3dgts) ± (2.5%+5dgts)
100mV
400V(<400V)
100mV
AC Voltage
750 V
± (1.0%+3dgts)
N/A
1V
400
µ
A
0.1
µ
A
4mA
1
µ
A
40mA
10
µ
A
400mA
± (2.5%+5dgts)
100
µ
A
4A
± (1.5%+3dgts)
1mA
AC Current
20A
± (2.0%+5dgts)
N/A
10mA