SCPI Command Reference
124
T200S
T200S Specific Commands
[SOURce]:WAVelength:SWEep:STOP
Description
This command sets the stop wavelength of a swept signal.
At *RST, this value will be set to MAXimum.
Syntax
[:SOURce]:WAVelength:SWEep:STOP<wsp><stop>[<unit>]|
MAXimum|MINimum|DEFault
Parameter(s)
³
stop
:
Value corresponding to the stop wavelength in meters (if no
unit is specified).
To know the valid wavelength range, use the query
[SOURce]:WAVelength:SWEep:STOP?
MAX parameters).
³
unit
:
Unit of the stop wavelength: M preceded by a possible
multiplier. Possible multipliers are listed in
³
MAXimum
Sets the stop wavelength to the maximum possible value.
³
MINimum
Sets the stop wavelength to the minimum possible value.
³
DEFault
Sets the stop wavelength to the default value.
To know the default wavelength, use the query
[SOURce]:WAVelength:SWEep:STOP?
parameter).
Example(s)
SOUR:WAV:SWE:STOP 1490 NM
SOUR:WAV:SWE:STOP? returns 1.490000E-006
See Also
[SOURce]:WAVelength[:CW|FIXed]
[SOURce]:WAVelength:SWEep:STARt