Test Results
Power Blazer
457
Alarms/Errors Overview
Section/Line / RS/MS
Alarms
LOF-S
(Loss Of Frame - Section) - SONET
RS-LOF
(Regeneration Section - Loss Of Frame) - SDH
RX
:
A SEF (SONET)/RS-OOF (SDH) defect on the incoming optical
signal persists for at least 3 milliseconds.
TX
: Generates non-valid framing bytes (A1 and A2).
SEF
(Severely Errored Framing) - SONET.
RS-OOF
(Regeneration Section - Out-Of-Frame) - SDH.
RX
: A minimum of four consecutive errored framing patterns are
received.
TX
: Generates four consecutive errored framing patterns.
TIM-S
(Trace Identifier Mismatch - Section) - SONET
RS-TIM
(Regeneration Section - Trace Identifier Mismatch) - SDH
RX
: The received J0 Trace doesn’t match the expected message value.
Only available when Enable TIM-S/RS-TIM check box is selected (refer
to
AIS-L
(Alarm Indication Signal - Line) - SONET
MS-AIS
(Multiplex Section - Alarm Indication Signal) - SDH
RX
:
Bits 6, 7 and 8 of the K2 byte contain the “111” pattern in five
consecutive frames.
TX
: Generates a SONET/SDH signal that contains a valid Section
Overhead (SOH) / Regenerator Section Overthead (RSOH) and an
all-ones pattern on the SPE.
Summary of Contents for Power Blazer 88200NGE
Page 1: ...www EXFO com User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE...
Page 14: ......
Page 20: ......
Page 64: ...Test Setup Test Applications 54 88000 Series OTN BERT...
Page 71: ...Test Setup Test Applications Power Blazer 61 SONET SDH BERT For SDH BERT...
Page 76: ...Test Setup Test Applications 66 88000 Series SONET SDH DSn PDH BERT For SDH...
Page 620: ......
Page 726: ......
Page 752: ......
Page 756: ......
Page 768: ......
Page 886: ...0 5 1 5 48 5 0 176 W d y K y K D...