Test Results
Power Blazer
441
Alarms/Errors Overview
OTUx
Note:
Available for OTU4, OTU3, OTU3e1, OTU3e2, OTU2, OTU2f, OTU1f, OTU2e,
OTU1e, and OTU1.
Alarms
AIS
(Alarm Indication Signal) available with serial interfaces only.
RX
: Polynomial number 11 (PN-11) is over all OTU frame bits including
FAS and MFAS for at least 3 consecutive 8192 bit-interval.
TX
: Generates polynomial number 11 (PN-11) over all OTU frame bits
including FAS and MFAS continuously.
BDI
(Backward Defect Indication)
RX
: The BDI bit in the SM overhead field (byte 3, bit 5) is “1” for at least
5 consecutive OTU frames.
TX
: Generates “1” for the BDI bit in the SM overhead field (byte 3, bit 5)
continuously.
BIAE
(Backward Incoming Alignment Error)
RX
: The BEI/BIAE bits in the SM overhead field (byte 3, bits 1 to 4) are
“1011” for at least 3 consecutive frames.
TX
: Generates “1011” for the BEI/BIAE bits in the SM overhead field
(byte 3, bits 1 to 4) continuously.
IAE
(Incoming Alignment Error)
RX
: The IAE bit in the SM overhead field (byte 3, bit 6) is “1” for at least
5 consecutive OTU frames.
TX
: Generates “1” for the IAE bit in the SM overhead field (byte 3, bit 6)
continuously.
Summary of Contents for Power Blazer 88200NGE
Page 1: ...www EXFO com User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE...
Page 14: ......
Page 20: ......
Page 64: ...Test Setup Test Applications 54 88000 Series OTN BERT...
Page 71: ...Test Setup Test Applications Power Blazer 61 SONET SDH BERT For SDH BERT...
Page 76: ...Test Setup Test Applications 66 88000 Series SONET SDH DSn PDH BERT For SDH...
Page 620: ......
Page 726: ......
Page 752: ......
Page 756: ......
Page 768: ......
Page 886: ...0 5 1 5 48 5 0 176 W d y K y K D...