Test Results
Power Blazer
361
Alarms/Errors Overview
Oversize
(available when the
Oversize Monitoring
check box is
selected.)
RX
: Frames larger than 1518 bytes
with a valid FCS; 1534 bytes
for
EoE; 1536 bytes
for PBB-TE.
Runt
RX
: Frames smaller than 64 bytes with an invalid FCS; 80 bytes for EoE;
82 bytes for PBB-TE.
Undersize
RX
: Frames smaller than 64 bytes with a valid FCS; 80 bytes for EoE; 82
bytes for PBB-TE.
Oversize Monitoring
check box, when selected (cleared by default),
allows monitoring the
Oversize
frame errors.
The following errors are only available with
Half Duplex
mode (only for
electrical interface at speeds of 10 Mbit/s and 100 Mbit/s).
Collision
RX
: Indicates the number of collisions on the link.
Late Coll.
RX
: Indicates the number of collisions that have occurred after a 64
bytes transmission.
Exc. Coll.
RX
: Indicates the number of frames that were sent 16 times
unsuccessfully due to consecutive collisions.