Summary Tabs
Ethernet and Fibre Channel Application
125
Test Summary
Clock Configuration
Clock Mode
indicates the clock mode selected during the test setup.
Available with 10Gig-E WAN interface only. Possible choices are:
Clock Mode
LAN WAN
Internal
: Internal clock of the unit (STRATUM 3). Internal is the only
available clock when 10GigE LAN transceiver mode is selected with
Frame
Analyzer
,
BERT
, or
RFC 2544
test.
X
X
External
: Clock from the connected
DS1/E1/2M
external clock signal (
For
IQS-8510G: EXT. CLK DS1/2M IN port; For IQS-8120NGE/IQS-8130NGE
and IQS-8525/IQS-8535: AUX
port). Once the test is created, refer to
Clock
Synchronization
-
to complete the external clock settings. For
IQS-8120NGE and IQS-8130NGE,
External
is not available in Dual Test Set
one-way latency mode.
-
X
Recovered
: Live clock from the optical port input signal involved in the test.
Recovered is the only available clock when 10GigE WAN transceiver mode is
selected with
Smart Loopback
test.
Recovered
is only available for Smart
Loopback test.
X
X
Backplane
: 8 kHz clock from another module on the IQS-600. For IQS-8535,
only available with 10Gig-E WAN interface. Note that the other module must
support the backplane clock feature and must be enabled (refer to
page 385 for more information).
-
X