Measuring Power
50
IQS-1700
Using the Offset Function
Using the Offset Function
The offset function is used when you want to take into account, in the
power displayed, a known gain or loss in the link that is not already
included in the signal reaching the detector. Contrary to the correction
factor, which applies to a specific wavelength, the offset value applies to
any wavelength when it is enabled in a specific channel.
To enter an offset value:
1.
From the
Settings
function tab, select the
Channels
tab.
2.
Select the channel for which you want to set the offset if you are using
a multichannel power meter.
3.
From the
Offset
list, use the arrow buttons to adjust the value.
4.
Click
Apply
to confirm your new setting.