Setting Up Your FTB-500
FTB-500
67
Recalibrating the Touchscreen
Recalibrating the Touchscreen
If you notice the touchscreen does not behave in the way it used to
(for example, it is now difficult to select items) it probably needs a
recalibration. You can perform a 4-point, 9-point (linearization), 16-point
(linearization), or even a 25-point (linearization) calibration. You can
perform a 25-point linearization when you need more accuracy on screen
edges and corners.
You can stop the calibration process at any time, but the touchscreen will
still need calibration. The parameters are taken into account only when the
process is complete.
Note:
If you have trouble accessing the touchscreen calibration feature because
the touchscreen is not behaving as expected, you can connect a USB
mouse.
I
MPORTANT
To get the optimum performance out of your touchscreen:
Always use the calibration tool provided with ToolBox (not the
tool provided with Windows).
Be as accurate as possible when you press the center of the
targets that appear during touchscreen calibration. This will
ensure a greater accuracy of the touchscreen, especially along
the edges and in corners.
Do not change the default resolution of the touchscreen
(800 x 600). Doing so will affect its performance and accuracy.