Test Results
376
FTB-700G/800 Series
Quality Level (SyncE)
Quality Level (SyncE)
The quality level characterizes the clock quality in terms of network
synchronization.
From the
Test
menu, tap
Results
, and the
Quality Level
tab.
Generated QL
indicates the Quality Level value that is generated.
Last Change
indicates the date and time of the Generated Quality
Level value has changed. For 1GE Electrical interface using
Local
Clock
set to
Slave
, the
Last Change
is not available.
Last QL Received
indicates the last Quality Level value received. The
pass/fail verdict is also displayed when both
QL Mismatch Monitoring
and
Pass/Fail Verdict
check boxes are selected.
Last Change
indicates the date and time of the Last QL Received has
changed.
QL Mismatch Monitoring
when selected (default), enables the Quality
Level mismatch monitoring.
Expected QL
allows the selection of the Expected Quality Level value.
Available when the
QL Mismatch Monitoring
check box is selected.
Quality Level
value
SSM
Description
QL-STU/UNK
0
Synchronized - Traceability Unknown
QL-PRS
1
Primary Reference Source Traceable (G.811)
QL-PRC
2
Primary Reference Clock Traceable (G.811)
QL-INV3
3
Quality Level Invalid 3
QL-SSU-A/TNC
4
Type I or V slave clock (G.812)
Traceable to Transit Node Clock (G.812 Type V)
QL-INV5
5
Quality Level Invalid 5
QL-INV6
6
Quality Level Invalid 6
QL-ST2
7
Traceable to Stratum 2 (G.812 Type II)
Summary of Contents for FTB-1
Page 1: ...User Guide FTB 700G 800 Series NetBlazer for FTB 1 ETHERNET AND MULTISERVICE TESTER...
Page 12: ......
Page 20: ......
Page 22: ......
Page 52: ......
Page 59: ...Test Setup Test Applications NetBlazer 47 SONET SDH BERT For SDH BERT...
Page 64: ...Test Setup Test Applications 52 FTB 700G 800 Series SONET SDH DSn PDH BERT For SDH...
Page 84: ......
Page 98: ......
Page 298: ......
Page 574: ......
Page 652: ......
Page 674: ...0 5 1 5 48 5 0 176 W d y K y K D...