Maintenance
38
Replacing Fuses
Replacing Fuses
The FLS-5834A CD/PMD Analyzer Source contains two F2.0L250V-type fuses
(IEC, 5 mm x 20 mm (0.197 in x 0.787 in), fast-acting, low breaking
capacity, 250 V). The fuse holder is located at the back of the CD/PMD
Analyzer Source, just beside the power inlet.
To replace a fuse:
1.
Turn off the unit and disconnect the power cord.
2.
Using a flat-head screwdriver as a lever, pull out the fuse holder.
3.
Verify and replace the fuses as necessary.
Fuse holder