Exergen TAT-2000C/SmartGlow Instruction Manual Download Page 1

Familiarize Yourself with the Instrument

To Scan: Depress the button. The instrument will continually 

scan for the highest temperature (peak) as long as the button is 

depressed. Beeping and LED flashing: Beep and LED flashing 

indicate a rise to a higher temperature, similar to a radar 

detector. Slow beeping indicates that the instrument is still 

scanning, but not finding any higher temperatures. Retain 

Reading: The reading will remain on the display for 30 seconds 

after the button is released. To Restart: Depress the button to 

restart. It is not necessary to wait until the display is clear, the 

thermometer will immediately begin a new scan each time the 

button is depressed

 and held depressed

.

 Quick Start Instructions

Your thermometer is the home version of the professional Temporal

Scanner 

used in thousands of hospitals and doctors’ offices, and is used the same 

way as doctors and nurses use the professional model.  As a technologically 

advanced instrument it must be used correctly.  Please follow the 

instructions below and read the user manual carefully.

Installing the Battery

1.

Locate the battery compartment at the back of the thermometer.

2.

Face the top of the thermometer toward you and place your thumb on the ridges.

3.

Press down firmly and slide the cover away from you, toward the bottom of the 

thermometer.

4.

Insert a 9-volt battery with the positive (small) terminal on the right. 

5.

Slide the battery compartment cover back into place.

Taking a Temperature

1.

Remove protective cap.

2.

Hold thermometer so that thumb or index finger is positioned over the

SCAN button (DO NOT depress until step four).

3.

Place the probe flush (flat) on the center of forehead.

4.

Press and hold the SCAN button.

5.

Lightly slide thermometer either left or right from the center of forehead to

the hairline, keeping the sensor flat and in contact with the skin for the

entire scan (2-3 seconds). You will hear beeping and a red light will flash

during the scan. This is normal, indicating that a measurement is taking place.

6.

Release the SCAN button and remove the thermometer from head once you 

reach hairline. 

7.

Read the temperature on display. 

If perspiration is present on forehead:

1.

Press and hold the SCAN button.

2.

Place the probe on the neck just behind the ear lobe.

3.

Release the SCAN button and remove the thermometer to get a reading.

Important Safety Instructions

READ ALLINSTRUCTIONS BEFORE USING

When using the product, especially when children are present, basic safety 

precautions should always be followed, including the following:

This product is intended for household use only. For information on thermometers for 

professional use, please see www.exergen.com, or call 617-923-9900.

Use this product only for its intended use as described in this manual.

Use of this product is not intended as a substitute for consultation with your physician. 

Do not take temperature over scar tissue, open sores or abrasions.

Basic safety precautions should always be observed, especially when this

product is used by, on or near children or invalids.

The operating environmental temperature range for this product is 60 to 104°F

(15.5 to 40°C).

Always store this thermometer in a clean, dry place where it will not become

excessively cold (-4°F/-20°C), or hot (122°F/50°C).

The thermometer is not shockproof. Do not drop it or expose it to shocks.

This thermometer is not intended to be sterile. Do not try to sterilize it. 

Follow the cleaning instructions as described in this manual.

Do not use this thermometer if it is not working properly, if it has been exposed to 

temperature extremes, damaged, been subject to electrical shocks or immersed in

 

water.

There are no parts that you can service yourself except for the battery, which you

should replace when low following the instructions in this manual. For service,

examination, repair, or adjustments, return your thermometer to Exergen.

Do not operate where aerosol spray products are being used or where oxygen is

being administered.

Do not take temperatures with this thermometer near places that are very hot,

such as fireplaces and stoves.

Do not use this thermometer outdoors.

Never drop or insert any object into any opening.

If your thermometer will not be used regularly, remove the battery to prevent

possible damage due to chemical leakage. If the battery leaks, remove carefully.

Do not allow bare skin to touch leaking fluid.

Dispose of used batteries properly. Do not wrap them in metal or aluminum foil.

Wrap them in newspaper before disposing of them. Do not burn them. Battery

may explode if overheated.

SmartGlow

: Eight Temperature Memory Mode

In memory mode, a small 

M   will appear on the display with memory location ( 1, 2, 3, 

4, 5, 6, 7, 8 ) and the memory location will alternate with the temperature stored in 

that location. The most recent temperature is stored in location 1 and the earliest 

temperature is stored in location 8. A display of 

188. 8   means that no temperature 

reading is stored in that memory location. Readings already in the memory when you

 

first use it are from the factory tests.

 

1.

To enter memory recall mode, start from a blank display, and

tap

 

the button, 

1 M  will appear and alternate with the

temperature

 

stored in that memory location.

2.

To access previously stored readings, tap the power button

again.

 

2 M will alternate with the temperature in that memory

location.

 

Repeat through 

8 M  to access all recent memory

readings.

3.

To exit memory recall mode and take a temperature, press

and

 

hold the button for 1-2 seconds until the display turns off.

Instructions for Use

Questions: 

Temporalscanner.com 

[email protected]

1-800-422-3006

Clinical studies, multiple 

language videos, & user 

manuals: exergen.com/s

Clinically proven

accuracy  for

all ages

LED Indicator Light

LED Indicator Light

Thumb-Push Ridges
Battery Compartment

9-volt  battery

SCAN Button

Automatic  turn-off in

30 seconds 

s

Protective Cap (remove during use) 

Probe Cone

Probe Sensor Lens

Probe

LCD Display Screen

SmartGlow

: Turn the Audible Beeper OFF or ON

Your Temporal

Scanner  is equipped with a beeper and front/back 

flashing LED’s which gives you a soft audible and visual 

feedback as you scan over the temporal artery on the forehead. 

The original factory setting is with the beeper and LED’s ON.

The TemporalScanner will automatically shut off 30 seconds after the last button 

press, from any mode of operation. The TemporalScanner must be OFF (blank 

screen) before taking a temperature. The TemporalScanner can be turned off 

from any programming mode by pressing and holding the button for 3 seconds.

When not in use: Thermometer will shut off automatically after 30 seconds.

To turn off immediately, press and release the SCAN button quickly. Always 

replace the cap to protect the sensor when not in use.

Scanning the Temporal Artery (TA)

Measure only the side of the head exposed to the environment.

  Anything 

covering the area to be measured (hair, hat, wig, bandages) would insulate the 

area, resulting in falsely high readings. Slide the thermometer straight across the 

forehead, not down the side of the face. Midline on the forehead, the TA is 

about a millimeter below the skin, whereas at the side of the face, the TA is 

much deeper, and measuring there would result in falsely low readings. When 

taking the temperature behind the ear (if there is perspiration on the forehead), first 

push away any hair, exposing the area.  Then, tuck the thermometer on the 

neck below the ear lobe, in the soft conical depression, (the place where 

perfume might be applied).  Wait about 60 seconds before measuring the same 

person again to avoid excessive cooling of the skin.  An infant is frequently 

swaddled in blankets and clothing covering the 

neck area. Unless visibly sweaty, 

one measurement at the TA area is typically all that is required. Should you feel 

the temperature is low, then push aside any clothing or blankets covering the 

neck area for ~30 seconds or so, and repeat the measurement on the neck 

behind the ear.  Factors that may affect measurement accuracy: The patented 

Arterial Heat Balance (AHB) technology in your TemporalScanner actually 

makes two separate measurements (1) the temperature of the skin over the 

temporal artery, and (2) the temperature of the room. To determine the most 

accurate reading, it measures skin temperatures some 1000 times a second as 

you sweep the TemporalScanner across the forehead. The AHB system then 

calculates how much the blood has cooled down during its journey from the 

heart to the skin over the temporal artery and makes allowance for this in the 

temperature it displays. The result is a highly accurate reading - delivered 

extremely fast and with no discomfort. 

Are there environmental effects?

As part of its Arterial Heat Balance system, the Temporal

Scanner  measures the 

temperature of the surrounding environment. For this measurement to be 

accurate, it needs to have become acclimatized to the temperature of the room 

in which it is to be used. If it is taken from a cold room into a hot room, or vice 

versa, allow it to acclimate for at least 30 minutes before using it. Avoid holding 

the Temporal

Scanner  by the head, as it will mistake the temperature of your 

hand for that of the room. 

If the beeper is on and you are concerned that the sound or flashing LED may wake a 

sleeping child (or adult), the beeper and front LED can be turned off as follows:

1.

 Starting with a blank display, tap the power button until the display flashes

loud,(tap 10 times, past the 8 memory locations and the F-C display)

2.

Press and hold the button until 

      

  S

hh

 

is displayed, indicating the quiet mode.

 

A

beep will indicate that the setting has been changed, and the

TemporaI

Scanner will automatically shut off.

3.

To turn the beeper and front LED back on, repeat step 1 to find 

Shh

 

 

then

 

step

2

 

to change into

 

 

loud.

SmartGlow

  Features - 

SmartGlow

  Model 

• Soft Glow Display: Soft lighting for easy reading in any light.

Perfect for checking

 

the temperature of a sleeping infant

(or adult) in a darkened room.

• Eight Temperature Memory: Automatically retains the

last 8 temperature

 

readings for instant recall to check on fever progression.

• Programmable Beeper: Beeping sound indicating scanning can be turned off

or

 

on to avoid waking a sleeping child (or adult). The flashing LED facing the

forehead while scanning turns off when the beeper is turned off.

• Programmable °F or °C: Select most convenient temperature scale.

• Elegant one-button programming and memory access.

Selecting 

°F

 

 

or 

°C

 

 

Mode  - 

SmartGlow

  Model  

1.

Starting from a blank display tap the button until the flashing

F - C  appears after the 8 memory locations (9 taps).  The current

temperature mode will appear in the upper right indicated by the

small 

°F  or °C.  The factory setting is °F. 

2.

To change from the 

°F   to the °C   mode, press and hold button

until the small 

°F   in the upper right of the display changes to °C.

 

A

beep will indicate that the setting has been changed, and the

TemporalScanner will automatically shut off.

3.

To change from the 

°C

 

 

to the 

°F

 

 

mode, repeat step 2, and

press and hold the button until the small 

°C

 

 

in the upper right of

the display changes to 

°F 

.

Save These Instructions