Calibration Verification Kit (CVK)

 

-for all Exergen TAT-5000 and 

TAT-5000S medical  thermometers

 

 

 

What is the  purpose of a  calibration  verification kit?

 

It allows the calibration of any Exergen thermometer to be verified in the 
hospital, on nursing floors, or in the field, conveniently, quickly, and 
accurately.

 

 

Why  is a CVK  necessary?

 

Calibration verification is a commonly required part of routine quality 
assurance programs, and also is used if a question is raised about the accuracy 
of a particular thermometer. Actual recalibration of an Exergen thermometer is 
never required unless it has been physically damaged or experiences 
component failure, in which case the calibration verification test identifies the 
problem device, which then is returned to the factory for repair.

 

 

What is in  the CVK?  

 

A special TAT-5000 Certified Master (CM) reference thermometer of the same 
calibration type (Arterial or Oral) in use by the institution, a portable hand-held 
reference blackbody, an AC plug-in adapter for long term use, a 9-volt battery 
for portable use, and a rugged storage case. 

  

Who uses  t h e   CVK?

 

The main users are biomedical/clinical engineering, blood donor and pheresis 
units (in the United States, blood donor and pheresis units are required by the 
American Association of Blood Banks to verify thermometers on a daily basis), 
and many nursing units where accurate temperature is critical such as 
hematology/oncology, and bone marrow and organ transplant units.

 

 

How  is the CVK ordered?

 

The order must include the serial number of any one of the TAT-5000 or TAT-
5000S thermometers that will be verified. This identification assures an exact 
calibration match to the CM reference thermometer.

 

 

Does the master reference thermometer need calibration? 

The calibration of the CM thermometer should be checked and factory re-

certified yearly. This is indicated by the due date on the calibration sticker 
affixed to the CM thermometer. On or before the due date the CM thermometer 
should be returned to Exergen for re-certification. The Exergen part number for 
re-certification is 139000, the cost is $75. 

 

 

 

Calibration Verification Kit 

P/N 129075 – Arterial 

P/N 129076 – Oral 

 

 

In  Use with TAT-5000

 

 

In  Use with TAT-5000S 

 

 
 

 

Handheld Blackbody

 

 

 

 

EXERGEN  

CORPORATION

 

400 Pleasant Street, Watertown, MA 02472

 

Phone 617

.

923

.

9900

 

        

Fax: 617-923-9911

 

www.exergen

.

com 

Email: [email protected]

 

Summary of Contents for Calibration Verification Kit

Page 1: ...uses the CVK The main users are biomedical clinical engineering blood donor and pheresis units in the United States blood donor and pheresis units are required by the American Association of Blood Banks to verify thermometers on a daily basis and many nursing units where accurate temperature is critical such as hematology oncology and bone marrow and organ transplant units How is the CVK ordered T...

Page 2: ...iomedical engineering and other areas requiring frequent accuracy verification of patient thermometry such as blood donor or pheresis units Calibration Verification Kit CVK The CVK includes a portable blackbody heat generator providing a stable source of heat in a small cavity This is used as a target reference to verify the calibration of any of Exergen s medical instruments against an Exergen Ce...

Page 3: ... please contact Exergen 2 Allow device to stabilize Once turned on allow approximately 5 minutes for warm up and stabilization time 3 Allow both the Certified Master and instruments to be tested to acclimate to the same ambient temperature Allow to equilibrate to room temperature for at least 10 minutes 4 Assure lens is clean Make sure the lens at the tip of the probe of all instruments including ...

Page 4: ...peat the process In the event they still differ by more than the acceptable limits call Exergen Corporation for replacement of the failed instrument 9 Heat Balance Awareness TAT 5000 and TAT 5000S models employ the patented arterial heat balance method to adjust to their surrounding ambient temperature for precise absolute accuracy If left sitting on the heat well they will assume the warmer tempe...

Page 5: ...re range 97 104 F 36 40 C may vary with battery voltage Cleaning wipe down with alcohol or cleaning solution do not immerse Recertification Certification is for one year Please call Exergen to arrange for recertification Figure 3 Power supply for PBB 1 If you have any questions about the calibration or operation of the Exergen Infrared Thermometers please email medical exergen com or call Exergen ...

Reviews: