METALJET D2+
Revision 19, November 2016
Page 6-5
START
Jet was stopped for
some reason
Restart source
Stable jet?
Section 5.8.4
Yes
End
Do you want to maximize
startup reliability?
No
Yes
Replace nozzle &
restart source
No
Replace nozzle &
restart source
Stable jet?
Section 5.8.4
Yes
End
No
First contact Excillum support. Replace liquid-
metal particle filter and then nozzle as
described in Section 6.2.5.
Claim warranty for nozzle and filter.
Stable jet?
Section 5.8.4
End
Yes
No
Contact Excillum
support for further
troubleshooting.
Figure 6-1.
Flowchart showing the decision process for nozzle and filter replacement.
Manufacturer information:
Follow recommendation above.
Service instructions:
Service manual under Section
Replace nozzle assembly
.