X-Cite

®

 110LED: Quick Start Guide

 

 

 

 

SIDE 1 

 

Warning 

– Refer to Safety Precautions Booklet 035-00636R for all safety precautions prior to use. 

User Guide 035-00575R download: 

http://www.excelitas.com/Pages/Product/X-Cite-110LED.aspx

 

 

 

WARNING:

 Never look at operating LED!  The light could severely damage the cornea and retina of the eye if the light is observed directly. Eye shielding 

must be used at all times as well as clothing to protect exposed skin  

Printed in Canada 

Refer to user guide for more details. 

035-00577R Rev. 2 

 

SETUP 
Unpack 

1.

 

Carefully unpack the unit and accessories from shipping carton.   

2.

 

When unpacking the light guide take care not to bend sharply. 

 

Install the Microscope Adaptor (if applicable) 

1.

 

Remove protective wrap from adaptor.   

2.

 

Insert the flange portion of the Microscope Adaptor into the light port on 

the microscope, and secure it using the hardware provided on the 

microscope.     

 

Connections and Positioning 

1.

 

Remove red caps from light guide.  Insert large end into light guide port on 

unit (up to “B” line). Insert small end into microscope light guide port (or 

adaptor), and secure it with hardware provided.  

2.

 

Position the equipment.  Ensure adequate clearance for ventilation around 

unit, and slack in the light guide. 

3.

 

Connect speedDIAL cable to “remote” port on main unit.   

4.

 

Connect USB cable to ports on main unit and computer.  (For Windows 

7/8.1, USB driver will install via internet after unit powers up, see user 

guide for additional information.) 

5.

 

Connect external power supply to main unit. 

6.

 

Connect AC power to properly grounded outlet.   

 

MANUAL OPERATION 

1.

 

Flip rocker switch to turn unit on.   

2.

 

When display shows “x%”, it is ready to use.  

3.

 

Click speedDIAL to illuminate a specimen. 

4.

 

Turn speedDIAL to adjust intensity. 

 
REGULATORY

 

(refer to user guide for complete details)

 

 

Product Safety and Electromagnetic Compatibility 

 

The X-Cite 110LED has been tested and found to comply with product safety and 

electromagnetic compatibility requirements.  For a complete list of tests and for 
certification details, please contact your X-Cite representative or visit 

http://www.excelitas.com/Pages/Product/X-Cite-110LED.aspx

.   

 

CE Marking 

 

 

Council Directive 2014/35/EU 

  Low Voltage Directive 

 

Council Directive 2014/30/EU   

  EMC Directive 

 

Council Directive 2012/19/EU 

  WEEE Directive 

 

Council Directive 2011/65/EU 

  RoHS 

 

 

 

BASIC TROUBLESHOOTING

 

(refer to user guide for complete details)

  

 

Category 

Symptom 

Action 

Error 

messages 

Error 1 (High Temp.) 

Turn off, let cool; check 

clearance around vents 

Error 3 or 5 (Internal) 

Turn off, restart   

Error 6 (Low Temp.) 

Allow unit to warm to room 

temperature and restart 

Intensity 

Low output intensity 

Check speedDIAL settings 

Check light guide position 

Check adaptor type 
Check microscope beam path, 

filters, apertures, shutters, 

objectives  

speedDIAL 

Double-click turns off LED  

Increase delay in the speed 

(Spd) setting 

New settings forgotten when 

unit is shut down  

After changing settings, wait at 

least five minutes before 

shutting unit down 

Dial does not scroll through 

the menu  

Turn dial in the opposite 

direction 

“Lock” icon on display, 

speedDIAL not responding to 

manual control 

Send “unlock” command via 

computer OR power down unit 

and restart 

TTL 

TTL response time is longer 

than usual 

Enable TTL mode 

Verify that timeout settings are 

appropriately set   

 

 

        

                              

                    

                   

 

WEEE Directive       

   China RoHS                  RoHS   

       FCC 

 