MFC
Functions
4.
67
Membrane sensor for
total chlorine TC1 and TC1-S
The membrane sensor measures the total amount of free and
combined chlorine (chloramine). The sensor demonstrates a slight
pH value dependency (see "Technical data“) and is therefore very
suitable for water with various pH values. The sensor may only be
used in clear water of potable water quality.
For applications in brine with a conductivity of 2.5–60 mS/cm
(approx. 4 % NaCl), an identically constructed sensor TC1-S must
be used with special electrolyte.
Commissioning
Caution!
The electrode fingers (D) and membranes (G) are extremely
sensitive! Do not touch, soil or damage! Before unscrewing the
filled membrane cap (E), push the elastomer seal (H) to one side
to permit the inflow of air through the vent (F) underneath it (at the
cap label)! Otherwise, the membrane (G) may be damaged due to
the development of underpressure. Do not remove the light
yellowish-gray deposit of the reference electrode (K) or wipe it in
the direction of the gold working electrode (J)!
Flush eyes and skin immediately with water after contact. Rinse
away spilled electrolyte with water.
1
Screw off the membrane cap (E) from the electrode shaft (B)
and fill with the included gel electrolyte up to the top brim. With
TC1-S a special brine electrolyte is used instead of the
standard electrolyte contained in the shipment.
2
Rub gold working electrode (J) with the included lapping paper
(special emery). To do this, lay the lapping paper on a paper
towel, take hold of a corner, and using the tip of the vertically-
held electrode shaft (B), slide it once or twice over the rough
side of the lapping paper.
3
Check whether the elastomer seal (H) completely closes the
valve opening (F).
4
Then screw the filled membrane cap (E) slowly (by hand) back
onto the electrode shaft (B).
Please note
As the excess electrolyte escapes through the valve opening (F)
under the elastomer seal (H), do not clamp it shut and do not press
it onto the elastomer seal (H).
5
Rinse off the escaped electrolyte with water.
Please note
The membrane cap (E) must be completely screwed (hand-tight)
onto the electrode shaft (B), so that no gap remains between the
two! After a run-in period of about one hour, the membrane sensor
is sufficiently run-in for an initial calibration to take place.
Calibration should be repeated after one day.
A A B 1 4 2 3
A
D
E
G
H
J
K
B
F
C
I
W3T171787
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...