64
WT.050.580.000.DE.IM.0714
Functions
MFC
4.
Membrane sensor for ozone
OZ7
The OZ7 membrane sensor enables the ozone content in all types
of water to be determined selectively, also in high-purity water (LF
> 1µS/cm), without cross-sensitivity to chlorine, bromine and
hydrogen peroxide, but with cross-sensitivity to chlorine dioxide
and peracetic acid. The sensor demonstrates no pH value
dependency (see "Technical data“) and is therefore very suitable
for water with various pH values.
Commissioning
Caution!
The electrode fingers (D) and membranes (G) are extremely
sensitive! Do not touch, soil or damage! Before unscrewing the
filled membrane cap (E), push the elastomer seal (H) to one side
to permit the inflow of air through the vent (F) underneath it (at the
cap label)! Otherwise, the membrane (G) may be damaged due to
the development of underpressure. Do not remove the light
yellowish-gray deposit of the reference electrode (K) or wipe it in
the direction of the gold working electrode (J)!
Flush eyes and skin immediately with water after contact. Rinse
away spilled electrolyte with water.
1
Screw off the membrane cap (E) from the electrode shaft (B)
and fill with the included gel electrolyte up to the top brim.
2
Rub gold working electrode (J) with the included lapping paper
(special emery). To do this, lay the lapping paper on a paper
towel, take hold of a corner, and using the tip of the vertically-
held electrode shaft (B), slide it once or twice over the rough
side of the lapping paper.
3
Check whether the elastomer seal (H) completely closes the
valve opening (F).
4
Then screw the filled membrane cap (E) slowly (by hand) back
onto the electrode shaft (B).
Please note
As the excess electrolyte escapes through the valve opening (F)
under the elastomer seal (H), do not clamp it shut and do not press
it onto the elastomer seal (H).
5
Rinse off the escaped electrolyte with water.
Please note
The membrane cap (E) must be completely screwed (hand-tight)
onto the electrode shaft (B), so that no gap remains between the
two! When the sensors are first commissioned, it is necessary to
operate the sensors for two or three hours in water with an ozone
content of >0.2 mg/l. The sensor needs this time to activate itself.
Following a run-in time of approx. three hours, a first calibration
can take place. Calibration should be repeated after one day.
A
C
I
G
H
J
K
B
F
E
D
W3T164494
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...