154
WT.050.580.000.DE.IM.0714
Operation
MFC
5.
5
Wait until the displayed value or the µA sensor signal no
longer changes for at least one minute..
6
Press the "ENTER" softkey to set the display to zero and
press "Enter" again to save the value.
Zero point calibration with
disinfectant-free water
7
If disinfectant-free water (e.g. by switching off the dosing
system) is available, zero point calibration can be performed
with it. To do this, switch off the dosing system and perform
steps 1, 2, 3, 5 and 6. It is necessary to observe the sample
water take-off and dosing delay time here!
8
Open the ball valve on the sample water inlet.
Measurement value calibration
(DPD)
9
After zero point calibration, wait at least 2 minutes.
10
Open the knurled nut on the drain by approximately one turn
and extract a specimen of the sample water.
11
Determine the free chlorine, ozone, chlorine dioxide or
potassium permanganate content, for example, with a
photometer.
12
Select the parameter "Cal. Span" and confirm with "Enter".
13
Enter the determined value with the arrow keys and then press
the "Enter" key to save.
This concludes the calibration.
Please note
In the DPD calibration of the Cl2++ measurement, the calibration
value should be greater than or equal to 25% of the measurement
range.
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...