DEPOLOX
®
5 E flow cell
Installation
4.
39
4.4.8
Connecting the sample water outlet
Please note
Do not install water carrying lines made of copper. These would falsify
the measurements.
Sample water outlet on non-pressurized version
Proceed as follows:
1
On the non-pressurized version, no back-pressure is permitted in the
cell body.
2
The sample water outlet must be open. We recommend using a fun
-
nel above the outlet.
3
The sample water outlet must be laid in such a way as to prevent
siphoning.
Image 4 Cross-section of DEPOLOX
®
5 E flow cell, non-pressurized
A Sample water inlet with shut-off ball valve
B Sample taking unit (drain on the drain screw)
C Sample water outlet
Sample water outlet on pressurized version
Proceed as follows:
1
On the pressurized version, a maximum back-pressure of 1.5 bar is
permitted on the sample water outlet.
2
Ensure that the drain screw (sample taking unit) is always closed.
A B C
Summary of Contents for WALLACE & TIERNAN DEPOLOX 5 E 700 P Series
Page 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Page 56: ...56 WT 050 810 040 DE IM 0519 Maintenance DEPOLOX 5 E flow cell 5...
Page 72: ...72 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 76: ...76 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 80: ...80 WT 050 810 040 DE IM 0519 Index DEPOLOX 5 E flow cell 8...